User's Manual
Table Of Contents
- About This Manual
- About the Clarius Ultrasound Scanner
- Using the Clarius Ultrasound Scanner
- Accessories
- Cleaning & Disinfecting
- Safety
- References
- Measurement Accuracy Tables
- Acoustic Output Tables
- Clarius Scanner C3 HD3: B-Mode
- Clarius Scanner C3 HD3: Color Doppler Mode
- Clarius Scanner C3 HD3: M-Mode
- Clarius Scanner C3 HD3: PW Doppler Mode
- Clarius Scanner C7 HD3: B-Mode
- Clarius Scanner C7 HD3: Color Doppler Mode
- Clarius Scanner C7 HD3: M-Mode
- Clarius Scanner C7 HD3: PW Doppler Mode
- Clarius Scanner EC7 HD3: B-Mode
- Clarius Scanner EC7 HD3: Color Doppler Mode
- Clarius Scanner EC7 HD3: M-Mode
- Clarius Scanner EC7 HD3: PW Doppler Mode
- Clarius Scanner L7 HD3: B-Mode
- Clarius Scanner L7 HD3: Color Doppler Mode
- Clarius Scanner L7 HD3: M-Mode
- Clarius Scanner L7 HD3: Needle Enhance B-Mode
- Clarius Scanner L7 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L7 HD3: PW Doppler Mode
- Clarius Scanner L15 HD3: B-Mode
- Clarius Scanner L15 HD3: Color Doppler Mode
- Clarius Scanner L15 HD3: M-Mode
- Clarius Scanner L15 HD3: Needle Enhance B-Mode
- Clarius Scanner L15 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L15 HD3: PW Doppler Mode
- Clarius Scanner L20 HD3: B-Mode
- Clarius Scanner L20 HD3: Color Doppler Mode
- Clarius Scanner L20 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L20 HD3: M-Mode
- Clarius Scanner L20 HD3: Needle Enhance B-Mode
- Clarius Scanner L20 HD3: PW Doppler Mode
- Clarius Scanner PA HD3: B-Mode
- Clarius Scanner PA HD3: Color Doppler Mode
- Clarius Scanner PA HD3: M-Mode
- Clarius Scanner PA HD3: PW Doppler Mode
- Clarius Scanner PA HD3: Transcranial B-Mode
- Clarius Scanner PA HD3: Transcranial Color Doppler Mode
- Clarius Scanner PA HD3: Transcranial M-Mode
- Clarius Scanner PA HD3: Transcranial PW Doppler Mode
- Revision History
Clarius Ultrasound Scanner - HD3 Scanners Safety Topics
revision 1 53
values associated with each mode and focal zone enabled. The system will return to
the previously selected state if a mode is turned off and then reselected.
• Scanner:
Each scanner type has unique specifications for contact area, beam shape, and center
frequency. Selecting a scanner initializes its default settings, which varies by scanner,
application, and selected mode. These defaults are set below the FDA limits for
intended use.
Example of reducing output:
Imagine we are getting ready to do a liver scan. The first thing we need to do is select the
appropriate scanner frequency. Next, we adjust the output intensity (or power) transmit
setting. We check to make sure that it is positioned at the lowest possible setting to produce
an image. We adjust the focus to the area of interest and then increase the receiver gain to
produce a uniform representation of the tissue. If we can obtain a good image by increasing
the gain, we can lower the output and continue to increase the gain. Only after making these
adjustments and if tissue penetration or echo amplitude levels are inadequate should we
increase the output to the next higher level.
Acoustics
The scanner is the most important factor in image quality. Optimal imaging cannot be obtained
without the correct scanner. The system is optimized for use based on your scanner selection.
The system limits patient contact temperature to 43°C (109°F), and acoustic output values to
their respective U.S. Food and Drug Administration limits. A power-protection circuit protects
against over-current conditions. If the power monitor protection circuit senses an over-current
condition, then the drive voltage to the scanner is shut off immediately, preventing
overheating of the scanner surface and limiting acoustic output. Validation of the power
protection circuit is done under normal system operation.
A temperature elevation of less than 1.5°C (2.7°F) is considered harmless to human tissue
(including embryo or fetus). Temperatures in excess of this may cause harm, depending on the
length of time maintained. A temperature elevation of 4°C (7.2°F), maintained for five minutes
or more, is considered to be potentially hazardous to a fetus or embryo.
Acoustic Artifacts
An acoustic artifact is information, present or absent in an image, which does not properly
indicate the structure or flow being imaged. Examples of acoustic artifacts that hinder proper
interpretation:
• Added objects displayed as speckle, section thickness, reverberation, mirror image, comet
tail, or ring down.
• Missing objects due to poor resolution.
• Incorrect object brightness due to shadowing or enhancement.