User's Manual
Table Of Contents
- About This Manual
- About the Clarius Ultrasound Scanner
- Using the Clarius Ultrasound Scanner
- Accessories
- Cleaning & Disinfecting
- Safety
- References
- Measurement Accuracy Tables
- Acoustic Output Tables
- Clarius Scanner C3 HD3: B-Mode
- Clarius Scanner C3 HD3: Color Doppler Mode
- Clarius Scanner C3 HD3: M-Mode
- Clarius Scanner C3 HD3: PW Doppler Mode
- Clarius Scanner C7 HD3: B-Mode
- Clarius Scanner C7 HD3: Color Doppler Mode
- Clarius Scanner C7 HD3: M-Mode
- Clarius Scanner C7 HD3: PW Doppler Mode
- Clarius Scanner EC7 HD3: B-Mode
- Clarius Scanner EC7 HD3: Color Doppler Mode
- Clarius Scanner EC7 HD3: M-Mode
- Clarius Scanner EC7 HD3: PW Doppler Mode
- Clarius Scanner L7 HD3: B-Mode
- Clarius Scanner L7 HD3: Color Doppler Mode
- Clarius Scanner L7 HD3: M-Mode
- Clarius Scanner L7 HD3: Needle Enhance B-Mode
- Clarius Scanner L7 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L7 HD3: PW Doppler Mode
- Clarius Scanner L15 HD3: B-Mode
- Clarius Scanner L15 HD3: Color Doppler Mode
- Clarius Scanner L15 HD3: M-Mode
- Clarius Scanner L15 HD3: Needle Enhance B-Mode
- Clarius Scanner L15 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L15 HD3: PW Doppler Mode
- Clarius Scanner L20 HD3: B-Mode
- Clarius Scanner L20 HD3: Color Doppler Mode
- Clarius Scanner L20 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L20 HD3: M-Mode
- Clarius Scanner L20 HD3: Needle Enhance B-Mode
- Clarius Scanner L20 HD3: PW Doppler Mode
- Clarius Scanner PA HD3: B-Mode
- Clarius Scanner PA HD3: Color Doppler Mode
- Clarius Scanner PA HD3: M-Mode
- Clarius Scanner PA HD3: PW Doppler Mode
- Clarius Scanner PA HD3: Transcranial B-Mode
- Clarius Scanner PA HD3: Transcranial Color Doppler Mode
- Clarius Scanner PA HD3: Transcranial M-Mode
- Clarius Scanner PA HD3: Transcranial PW Doppler Mode
- Revision History
Clarius Ultrasound Scanner - HD3 Scanners Spaulding Classification
revision 1 37
Because the Fan cannot be submerged in liquid, never use high-level disinfectants. Always use
intermediate-level disinfectants. For a list of disinfectants recommended for intermediate
disinfection of the Fan, refer to Cleaners & Disinfectants on page 70.
To disinfect the Fan:
1. Make sure that the fan is detached from the scanner.
2. Disinfect the fan by wiping with a cloth moistened with a compatible disinfectant.
Alternatively, use a premoistened disinfectant wipe.
3. Air-dry. Alternatively, use a non-linting cloth.
4. Examine the fan for damage, such as cracks or splitting. If damage is evident, do not
use the fan and contact Clarius Mobile Health.
Disinfecting the Charger
Before you begin disinfecting, make sure you have cleaned the charger (see Cleaning the
Charger on page 35).
Because the Charger cannot be submerged in liquid, never use high-level disinfectants. Always
use intermediate-level disinfectants. For a list of disinfectants recommended for intermediate
disinfection of the Charger, refer to Cleaners & Disinfectants on page 70.
To disinfect the Charger:
1. Disinfect the charger by wiping with a cloth moistened with a compatible disinfectant.
Alternatively, use a premoistened disinfectant wipe.
2. Air-dry. Alternatively, use a non-linting cloth.
3. Examine for damages, such as cracks or splits. If damage is evident, do not use the
charger and contact Clarius Mobile Health.
Spaulding Classification
The level of cleaning and disinfecting required for your Clarius Scanner HD3 is based on the
Spaulding classification system. Following the correct classification will help reduce cross-
contamination and infection.
Each Spaulding classification mandates a specific level of cleaning and disinfecting of the
equipment before it can be used in the next exam. Determine the Spaulding classification
based on your scanner’s usage.