User's Manual
Table Of Contents
- About This Manual
- About the Clarius Ultrasound Scanner
- Using the Clarius Ultrasound Scanner
- Accessories
- Cleaning & Disinfecting
- Safety
- References
- Measurement Accuracy Tables
- Acoustic Output Tables
- Clarius Scanner C3 HD3: B-Mode
- Clarius Scanner C3 HD3: Color Doppler Mode
- Clarius Scanner C3 HD3: M-Mode
- Clarius Scanner C3 HD3: PW Doppler Mode
- Clarius Scanner C7 HD3: B-Mode
- Clarius Scanner C7 HD3: Color Doppler Mode
- Clarius Scanner C7 HD3: M-Mode
- Clarius Scanner C7 HD3: PW Doppler Mode
- Clarius Scanner EC7 HD3: B-Mode
- Clarius Scanner EC7 HD3: Color Doppler Mode
- Clarius Scanner EC7 HD3: M-Mode
- Clarius Scanner EC7 HD3: PW Doppler Mode
- Clarius Scanner L7 HD3: B-Mode
- Clarius Scanner L7 HD3: Color Doppler Mode
- Clarius Scanner L7 HD3: M-Mode
- Clarius Scanner L7 HD3: Needle Enhance B-Mode
- Clarius Scanner L7 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L7 HD3: PW Doppler Mode
- Clarius Scanner L15 HD3: B-Mode
- Clarius Scanner L15 HD3: Color Doppler Mode
- Clarius Scanner L15 HD3: M-Mode
- Clarius Scanner L15 HD3: Needle Enhance B-Mode
- Clarius Scanner L15 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L15 HD3: PW Doppler Mode
- Clarius Scanner L20 HD3: B-Mode
- Clarius Scanner L20 HD3: Color Doppler Mode
- Clarius Scanner L20 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L20 HD3: M-Mode
- Clarius Scanner L20 HD3: Needle Enhance B-Mode
- Clarius Scanner L20 HD3: PW Doppler Mode
- Clarius Scanner PA HD3: B-Mode
- Clarius Scanner PA HD3: Color Doppler Mode
- Clarius Scanner PA HD3: M-Mode
- Clarius Scanner PA HD3: PW Doppler Mode
- Clarius Scanner PA HD3: Transcranial B-Mode
- Clarius Scanner PA HD3: Transcranial Color Doppler Mode
- Clarius Scanner PA HD3: Transcranial M-Mode
- Clarius Scanner PA HD3: Transcranial PW Doppler Mode
- Revision History
Clarius Ultrasound Scanner - HD3 Scanners Disinfecting
revision 1 35
Cleaning the Fan
To clean the Fan:
1. Remove the fan from the Clarius Scanner HD3.
2. Wipe down all surfaces using a premoistened disinfectant wipe.
For a list of compatible cleaners, see Cleaners & Disinfectants on page 70.
3. Repeat with new cleaning material if necessary.
4. Air-dry the fan. Alternatively, use a non-linting cloth.
When you are done, keep the two components separate. You will be disinfecting them
individually.
Cleaning the Charger
To clean the Charger:
1. Unplug the Charger from the power source.
2. Disconnect the Micro USB cable from the Charger.
3. Wipe down all surfaces using a premoistened disinfectant wipe. Do not submerge the
Charger in any liquid.
For a list of compatible cleaners, see Cleaners & Disinfectants on page 70.
4. Repeat with new cleaning material if necessary.
5. Air-dry the Charger. Alternatively, use a non-linting cloth.
Disinfecting
Disinfecting the Clarius Scanner HD3
Before you begin disinfecting, make sure you have cleaned the scanner (see Cleaning the
Clarius Scanner HD3 on page 34).
Disinfecting requires that you choose the proper disinfecting level. Determine the necessary
disinfection level by referring to Spaulding Classification on page 37. Once you have
determined the required disinfecting level, have the disinfectant ready and follow one of the
appropriate procedures below. Note that different levels of disinfection require different steps,
not just different solutions.