Cleaning the Fan
To clean the Fan:
1. Remove the fan from the Clarius Scanner HD3.
2. Wipe down all surfaces using a premoistened disinfectant wipe.
For a list of compatible cleaners, see Cleaners & Disinfectants on page 70.
3. Repeat with new cleaning material if necessary.
4. Air-dry the fan. Alternatively, use a non-linting cloth.
When you are done, keep the two components separate. You will be disinfecting them
individually.
Cleaning the Charger
To clean the Charger:
1. Unplug the Charger from the power source.
2. Disconnect the Micro USB cable from the Charger.
3. Wipe down all surfaces using a premoistened disinfectant wipe. Do not submerge the
Charger in any liquid.
For a list of compatible cleaners, see Cleaners & Disinfectants on page 70.
4. Repeat with new cleaning material if necessary.
5. Air-dry the Charger. Alternatively, use a non-linting cloth.
Disinfecting
Disinfecting the Clarius Scanner HD3
Before you begin disinfecting, make sure you have cleaned the scanner (see Cleaning the
Clarius Scanner HD3 on page 34).
Disinfecting requires that you choose the proper disinfecting level. Determine the necessary
disinfection level by referring to Spaulding Classification on page 37. Once you have
determined the required disinfecting level, have the disinfectant ready and follow one of the
appropriate procedures below. Note that different levels of disinfection require different steps,
not just different solutions.