User's Manual
Table Of Contents
- About This Manual
- About the Clarius Ultrasound Scanner
- Using the Clarius Ultrasound Scanner
- Accessories
- Cleaning & Disinfecting
- Safety
- References
- Measurement Accuracy Tables
- Acoustic Output Tables
- Clarius Scanner C3 HD3: B-Mode
- Clarius Scanner C3 HD3: Color Doppler Mode
- Clarius Scanner C3 HD3: M-Mode
- Clarius Scanner C3 HD3: PW Doppler Mode
- Clarius Scanner C7 HD3: B-Mode
- Clarius Scanner C7 HD3: Color Doppler Mode
- Clarius Scanner C7 HD3: M-Mode
- Clarius Scanner C7 HD3: PW Doppler Mode
- Clarius Scanner EC7 HD3: B-Mode
- Clarius Scanner EC7 HD3: Color Doppler Mode
- Clarius Scanner EC7 HD3: M-Mode
- Clarius Scanner EC7 HD3: PW Doppler Mode
- Clarius Scanner L7 HD3: B-Mode
- Clarius Scanner L7 HD3: Color Doppler Mode
- Clarius Scanner L7 HD3: M-Mode
- Clarius Scanner L7 HD3: Needle Enhance B-Mode
- Clarius Scanner L7 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L7 HD3: PW Doppler Mode
- Clarius Scanner L15 HD3: B-Mode
- Clarius Scanner L15 HD3: Color Doppler Mode
- Clarius Scanner L15 HD3: M-Mode
- Clarius Scanner L15 HD3: Needle Enhance B-Mode
- Clarius Scanner L15 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L15 HD3: PW Doppler Mode
- Clarius Scanner L20 HD3: B-Mode
- Clarius Scanner L20 HD3: Color Doppler Mode
- Clarius Scanner L20 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L20 HD3: M-Mode
- Clarius Scanner L20 HD3: Needle Enhance B-Mode
- Clarius Scanner L20 HD3: PW Doppler Mode
- Clarius Scanner PA HD3: B-Mode
- Clarius Scanner PA HD3: Color Doppler Mode
- Clarius Scanner PA HD3: M-Mode
- Clarius Scanner PA HD3: PW Doppler Mode
- Clarius Scanner PA HD3: Transcranial B-Mode
- Clarius Scanner PA HD3: Transcranial Color Doppler Mode
- Clarius Scanner PA HD3: Transcranial M-Mode
- Clarius Scanner PA HD3: Transcranial PW Doppler Mode
- Revision History
Clarius Ultrasound Scanner - HD3 Scanners Imaging
revision 1 24
1. After three seconds, it decreases frame rate.
2. After 30 seconds of decreased frame rate, it freezes.
3. After 10 seconds in freeze mode, it goes idle.
4. After 15 minutes of idle time, it shuts down
When you go to the image acquisition page to begin an exam, the Clarius Scanner HD3
automatically switches from standby mode to scanning mode. For instructions on using the
imaging tools, go to support.clarius.com/hc/en-us.
A typical use of the Clarius Scanner HD3 is described as five continuous scanning minutes
followed by 10 minutes in standby mode (or turned off).
• Notifications and alerts from third-party applications may interrupt you or the Clarius App,
thereby interfering with the exam. Configure your smart device in accordance with your
institution's security policies.
• Vibration range that is too high for the scanner may cause the scanner to malfunction
during an exam.
• Using improper gel type or combining different gel types may expose patients to risks and
produce poor-quality images.
For proper transmission of the acoustic beam, use only Aquasonic 100, and use it only before
its expiry date. Download the usage instructions from parkerlabs.com and read all the
information before operating the device.
Do not use:
• Lotion-based products or gels that contain mineral oil.
• Hand-sanitizing gels.
• Scanners left soaking in gel.
Scanner Notifications
The Clarius Ultrasound Scanner displays no error messages. Instead, the Clarius Scanner HD3
presents visual notifications in the form of status lights, and audible notifications in the form of
status alerts.
When scanning for more than 15 consecutive minutes, with any Clarius Scanner HD3, the fan
must be used. For information on the fan, see Fan on page 31.