revision 1 17
Precautions
For Use in Surgical Environments
Before you use the Clarius Scanner HD3 for intra-operative procedures or in a surgical
environment, follow the instructions for high-level disinfection (for instructions see High-Level
Disinfection on page 36), then cover the Clarius Scanner HD3 with a sheath:
• Use only CIVCO REF 610-1212, Safersonic DAN-33803-WG, and Safersonic DAN-33803-NT.
• Follow the usage instructions provided by the manufacturer.
When you have finished using the Clarius Scanner HD3, immediately clean it (for instructions
see Cleaning the Clarius Scanner HD3 on page 34), followed by another high-level disinfection.
If the sheath breaks during the intra-operative procedure, dispose the sheath and follow the
same cleaning and high-level disinfecting process as above, then cover the Clarius Scanner
HD3 with a new sheath before continuing to use it.
For Use in Endocavitary Procedures
Before using the Clarius Scanner HD3 for endocavitary procedures (trans-vaginal and/or trans-
rectal), inspect the probe for any rough surfaces, sharp edges, or sharp corners that may cause
harm to the patient. Next, perform a high-level disinfection (for instructions see High-Level
Disinfection on page 36), then cover the Clarius Scanner EC7 HD3 with a model #2220 sheath
by Protek Medical Supplies Inc. Make sure the entire length of the end piece is protected.
When you have finished using the Clarius Scanner EC7 HD3, immediately clean it (for
instructions see Cleaning the Clarius Scanner HD3 on page 34), followed by another high-level
disinfection. If a Fan (CIDN 50-02-00077) was used, remove the fan prior to the high-level
disinfection and follow the instructions for an intermediate-level disinfection (for instructions
see Disinfecting the Fan on page 36).
If the sheath breaks during an endocavitary procedure, dispose the sheath and follow the
same cleaning and high-level disinfecting process as above, then cover the Clarius Scanner
HD3 with a new sheath before continuing use.
When using a biopsy guide, use one of the following recommended products:
• Civco Reusable Endocavity Needle Guide (Product Number 613-198)
• Civco Disposable Endocavity Needle Guide (Product Numbers 667-089 and 667-090)
Follow the usage instructions provided by the manufacturer.
Do not reuse biopsy guides unless instructed by the manufacturer.