User's Manual

Table Of Contents
Clarius Ultrasound Scanner - HD3 Scanners
revision 1 106
Clarius Scanner L20 HD3: Color Doppler Mode
Index Label MI TIS TIB TIC
At
Surface
Below
Surface
At
Surface
Below
Surface
Maximum Index Value 0.243 0.021 0.022 (a)
Index Component Value
0.021 0.021 0.022 0.021
Acoustic
Parameters
p
r,a
at Z
MI
(MPa) 0.927
P(mW) 0.298 0.298 #
P
1x1
(mW) 0.298 0.298
z
s
(cm) 1.00
z
b
(cm) 1.00
Z
MI
(cm) 1.00
Z
pii,a
(cm) 1.00
f
awf
(MHz) 14.6 14.6 14.6 #
Other Information prr (Hz) 2080
srr (Hz) 13.0
n
pps
12
I
pa,a
at Z
pii,a
(W/cm
2
)
32.4
I
spta,a
at Z
pii,a
or Z
sii,a
(mW/cm
2
)
0.492
I
spta,a
at Z
pii
or Z
sii
(mW/cm
2
)
1.34
p
r
at Z
pii
(MPa) 1.53
Operating Control
Conditions
Control 1 
Control 2
Control 3
Control 4
Control 5
Control 6
Control 7
(a) This scanner is not intended for transcranial or neonatal cephalic use.
# Not reported for this operating condition; the maximum index value is not reported for the reasons shown in the Maximum
Index Value fields.
Control 1: Exam Type: Vascular; Depth: 4 cm; Mode: CD