revision 1 96
Clarius Scanner L7 HD3: Needle Enhance B-Mode
Index Label MI TIS TIB TIC
At
Surface
Below
Surface
At
Surface
Below
Surface
Maximum Index Value 0.987 0.329 0.501 (a)
Index Component Value
0.329 0.329 0.501 0.329
Acoustic
Parameters
p
r,a
at Z
MI
(MPa) 2.26
P(mW) 17.6 17.6 #
P
1x1
(mW) 13.2 13.2
z
s
(cm) 2.00
z
b
(cm) 2.00
Z
MI
(cm) 2.00
Z
pii,a
(cm) 2.00
f
awf
(MHz) 5.24 5.24 5.24 #
Other Information prr (Hz) 4800
srr (Hz) 25.0
n
pps
2
I
pa,a
at Z
pii,a
(W/cm
2
)
258
I
spta,a
at Z
pii,a
or Z
sii,a
(mW/cm
2
)
11.5
I
spta,a
at Z
pii
or Z
sii
(mW/cm
2
)
23.6
p
r
at Z
pii
(MPa) 3.24
Operating Control
Conditions
Control 1 
Control 2
Control 3
Control 4
Control 5
Control 6
Control 7
(a) This scanner is not intended for transcranial or neonatal cephalic use.
# Not reported for this operating condition; the maximum index value is not reported for the reasons shown in the Maximum
Index Value fields.
Control 1: Exam Type: MSK; Depth: 4 cm; Mode: B