RetCam Shuttle User Manual and Service Instructions PN 20-000238 Rev. B
Maintenance and Support 59
Figure 43 Inverted lens piece showing permissible depth of immersion
Cleaning the Rest of the System
As with any  medical device,  use good public health  practices when handling  the
equipment, based on CDC guidelines. In addition, as with typical office equipment,
a  gentle  wiping  with  a  cloth  moistened  with  mild  soap  and/or  water  is
recommended.  Do not spray cleaning solutions directly on the  computer or other
modules, spray the cloth instead. The cart and side panels also cannot be sprayed,
but only the cloth with which you wipe, to avoid introducing spray or mist inside the
system. Use only a mild soap and/or water solution.
The lens piece should be checked before each use for damage to the front contact
lens  and  surrounding  surface.  The  interconnect  cable  and  connectors  should  be
checked for damage to the end connectors and protective covers.
WARNING: Always inspect the lens before use for damage such as chips, cracks or
roughness that could injure the patient’s eye.
Disposal of Materials
Dispose of waste materials according to local and national requirements. Contact
Clarity Service if additional assistance is required.
4 mm (0.16‚Ä