Product Manual

5
Note: Ensure the type of strip selected corresponds with the strip to be
used. If it does not, an error message will be displayed.
Mode
There are three test mode options. Press ENTER when Mode is selected to change the
Mode.
Note: The Mode can be changed during normal testing for STAT tests.
Routine Test
Use for normal urine testing. The default test number ranges from 1 to 99999. It resets
to 1 every day automatically if Auto Number Reset is set to Yes.
STAT Test
Use for emergency urine testing. The default test number is from 100001 to 199999. It
resets to 100001 each day automatically if Auto Number Reset is set to Yes. The
analyzer will automatically switch to STAT test and can not be changed if the QC test
has failed.
QC Test
Use to test positive/negative controls. If QC lockout is on, this test will be run
automatically. The test number ranges from 200001 to 299999. It resets to 200001
every day automatically if Auto Number Reset is Yes.
Note: Ensure QC Test mode is used for testing positive and negative
controls. Use of any other mode will report the testing results as a
regular specimen and will not display proper “Pass/Fail” QC test
results.
Test Number
Select Test Number and press ENTER.
Enter the new Test Number. Press ENTER to confirm. Press CANCEL to delete any
changes and keep the existing number.
The first digit in the number will depend on the Test Mode. The first digit will be 0 for
Routine, 1 for STAT and 2 for QC Test. This first digit is fixed and cannot be changed
from this screen. Tests run after this number is changed will be set sequentially from the
new number.