Place the strip with the reagent area facing up, onto the Strip Holder Channel. Make
sure the strip end touches the Strip Holder Backstop as shown below.
Note: In order to avoid contamination of the strip holder and/or the optical
sensor, make sure that the strip holder is fully extended and at a
standstill before placing or removing a urine strip from the strip
holder.
When the countdown clock on the display reaches 1, the Strip
Holder will carry the strip inside and begin testing.
An automatic calibration is done before each test.
The results will be displayed on the screen and stored in memory
after each test. Any abnormal results will be highlighted on the
screen and flagged on the print out.
If Auto-print is set to on the results will be printed. If Auto-print is
set to off, press Print to print the results.
Remove the used strip from the Strip Holder when the strip carrier moves out. Discard
the used strip according to local regulations. Repeat the previous steps to test more
urine.
Strip Test – Continuous Test Mode
Press START when the Strip Holder is fully extended to begin the test. The following
screen will be displayed. Wait for the audible triple beep to immerse the strip.
Note: Timing is set up to process a new strip every 30 seconds.
Once START is pressed, the screen will change to the following: