Product Manual

21
Once START is pressed, the screen below will show.
The countdown clock is displayed on the bottom right. The clock will start to count down
from 65. The operator has 3 seconds before the triple beep sounds to immerse the strip
into the urine.
Upon hearing the triple beep and/or seeing the countdown clock reach 62, completely
immerse the reagent areas of the strip in fresh, well-mixed urine. Immediately remove
the strip to avoid dissolving the reagents.
Run the edge of the strip against the rim of the urine specimen container to remove
excess urine. Hold the strip in a horizontal position. Bring the edge of the strip into
contact with an absorbent material (e.g. a paper towel). This prevents mixing chemicals
from adjacent reagent areas.