Product Manual
20
Note: Refer to Section 4, Analyzer Setup, to change the settings.
When there are only 5 strips remaining, will flash to prompt that a new canister of
strips is required. The number of strips left can be seen under .
When the current run is completed, enter the new canister code. The number under
will be increased accordingly.
Note: Refer to the following for more information on the Strip
Lockout function. Once a canister code is accepted, the analyzer
will automatically recognize the Type of Strip and the number of
strips per canister should be tested. The analyzer will keep track of
the number of strips available for testing and the strips remaining
will be displayed under the lock symbol. Continue testing as
usual until the analyzer prompts that a new canister code is
required. When the analyzer prompts that a new canister code
is required, scan or manually enter a canister code from a new
canister to be used, or from the one currently in use. The
analyzer will not accept canister codes that have already been
entered and will lockout the user until a new canister code is
entered. One canister code can be entered for each Type of Strip,
and will be managed independently of other strip types by the
analyzer.
Normal Operation, no Barcode reader
Sample/Strip Preparation
Allow the strip, urine specimen, and/or controls to reach room temperature at 15-30ºC
(59-86ºF) prior to testing.
Note: Only Clarity
Urinalysis Reagent Strips work with the Clarity
Urocheck 120C Urine Analyzer. Use of any other strips will cause
inaccurate results.
Remove the strips from the closed canister. Use them as soon as possible. Tightly
close the canister after removing the strips.
Strip Test – Single Test Mode
Press START for the strip prompt. Wait for the audible triple beep to immerse the strip.