INSTALLATION PROCEDURES
INITIAL SETUP
Connect the Reagents
The connections for the reagents are at the bottom right corner on the left side of the
instrument (Figure 3.2-1).
1. Connect the diluent reagent using blue stripe tubing (PN 3202209-0) to the blue fitting
(diluent). Use a Luer fitting (PN 6232503-8).
2. Connect the lytic reagent using EVA tubing (PN 3202221-9) to the white fitting (lyse).
Use a Luer fitting (PN 6232503-8). Use polyurethane tubing (PN 3202036-4) to couple
the EVA tubing to the Luer fitting and to the pickup tube.
3. Ensure that the lyse container is no higher than 6 in. above or lower than 36 in. below
the lyse pump and that the lyse tubing is no longer than 5 ft.
4. Connect the waste container using red stripe tubing (PN 3202205-7) to the red fitting
(waste). Use a Luer fitting (PN 6232503-8).
5. If a cubitainer is used for waste, connect the waste alarm cable (PN 6028669-8) between
the waste pickup tube and the waste sensor connector. The waste sensor connector is
located to the left of the white fitting (lyse).
Figure 3.2-1 Reagent Connections
7242016A