Service manual
PN 4237242C
2.5-6
INSTRUMENT DESCRIPTION
DATA ACQUISITION
Test Points
Figure 2.5-3 Vacuum Sensor Card
Sensor Processing Adapter with Diagnostics (SPAD) Card
The SPAD card is located in the upper chassis (Figure 2.2-1). Its primary function is to
acquire and process instrument data. The SPAD card is a full-size, AT-type card designed to
plug into a 16-bit slot in the AT motherboard.
The SPAD card acquires WBC, RBC, and Plt pulse data from the Sensor Preamp Adapter card.
This data is counted, edited, channelized, and converted to digital format. It is then made
available to the processing software via the AT bus interface.
Table 2.5-4 Vacuum Sensor Card Test Points
Test Point Description
TP1 8.006 V reference voltage (7.964 V to 8.049 V)
TP2 Transducer output
TP3 Gain adjustment output
TP4 Ground
TP5 VAC OUT
J3
(To
solenoid
LV2)
J2
(To
solenoid
LV1)
J1
(
+ 15 V in
VAC OUT)
7242014A
TEST
POINTS
(1-5)
TP1