Service manual

PN 4237242C
2.3-5
INSTRUMENT DESCRIPTION
FLUIDICS PANEL
2
Figure 2.3-3 Probe-Wipe Traverse Assembly
Peristaltic Pump Assembly
The Peristaltic Pump Assembly is located on the lower left side of the instrument and consists
of a plate holding four peristaltic pumps and their associated tubing. See Figure 2.3-4.
Table 2.3-2 gives each pump’s location and function. The green and blue filters are also found
on this assembly. The green filter, a 0.45-µ fluid barrier, is found in the air lines. It blocks
fluid that might get into the line, from going any further. The blue filter, a 10-µ filter, is in the
fluid lines. The blue filter blocks particulate and air bubbles.
Figure 2.3-4 Peristaltic Pump Assembly
PROBE
MOTOR
TRAVERSE
HOUSING
SAMPLE
ASPIRATE
PROBE
PROBE/
OPTO SENSOR
CARD
TRAVERSE
MOTOR
SPRING BELT
TENSIONER
7242006A
7242007A
PM1
(AIR/MIX)
PM2
(DILUENT)
PM3
(RINSE)
PM4
(WASTE)
GREEN
AIR FILTER
(0.45µ)
BLUE
FLUID FILTER
(10µ)