Service manual

PN 4237242C
2.1-4
INSTRUMENT DESCRIPTION
SYSTEM OVERVIEW
The MD II has seven diluter tables or cycles. These are the aspirate, power up, startup,
shutdown, prime, latex and clean bath tables. Chapter 6 has timing charts for all the diluter
tables. Because they are the most useful in describing and troubleshooting instrument
operation, the aspirate cycle and the power-up cycle are described in detail. Make sure that
you use the most recent timing chart when troubleshooting timing concerns. Table 2.1-1
gives a list of the functions that can be used by a diluter table.
Table 2.1-1 Functions the Diluter Table Can Use
Motor Functions Solenoids & Valves S/W Requests
Asp Syringe Fill Vac Xdcr Vent Read Hgb Sample
Asp Syringe Disp Vac Chmbr Vent Read Hgb Blank #1
Dil Srng Fil Vac Select Read Hgb Blank #2
Dil Syringe Disp Sweep Flow WBC/RBC/Plt
Probe Down RBC Aper Vac RBC/Plt DAQ
Probe Up WBC Aper Vac WBC DAQ
Probe Aspirate Bath Drain Sel Vacuum Offset
Probe WBC PRB Wipe Vac Aspirate Key Wait
Probe RBC Air Mix Select Delay
Mix Bubbles Waste Select Zap Red Aperture
Drain Bath Rinse Sel Zap Wht Aperture
Rinse Bath Mix Sel Chk Dil Snsr
Dil Rsvr Fil Prefill Select Wait for Enter
Dil Rsvr Overfill Dil Prefil/Dis Probe Warn
Dil Srng Vlv Dil Overfill
Dil Dispense Shutdown Message
Deliver Lyse Do CBC
Inc. Cycle # Do Diff
Vacuum Pump Print Results
No Selection Rpt Activity
C