Service manual
PN 4237242C
C.2-1
MESSAGE/ERROR CODE LISTINGS
INSTRUMENT FATAL ERROR MESSAGES
C
C.2 INSTRUMENT FATAL ERROR MESSAGES
Table C.2-1 Fatal Error Messages
Error Code Displayed Message Description
001
SPAD Board Failure
Currently not used.
002
URA Board Failure
Currently not used.
003
DRA Board Failure
A motor driver did not generate an interrupt to indicate that it finished an
operation.
004
SPAD Board Failure
While channelizing pulses, a buffer overload occurred.
005
System Disk File Corrupt
The system was unable to load a resource file from the Program Disk.
006
Insufficient RAM
The system attempted to allocate memory to perform a function and there
was insufficient memory available.
008
URA Board Failure
The analog multiplexer on the URA card failed.
009
URA Board Failure
There was a timeout while the URA card was performing an A/D conversion.
010
Keypad Failure
Currently not used.
011
URA CMOS Failure
There is a problem with the data stored in the URA CMOS.
012
System Disk File Corrupt
The system could not load an acceptable PD.DAT file from the Program
Disk.
013
CPU Fatal Error
The CPU signaled the system software of a problem. This generally means it
tried to execute an illegal instruction.
014
Power Supply Failure
There was a failure from a voltage check of the +15, +5, +12 and supplies.
015
Power Supply Failure
There was a failure when checking the +24 V power supply.
016
System Disk File Corrupt
The system was unable to load a good printer report template file from the
Program Disk.
017
Unable to Sense Diluent Level
During self-test, the system was unable to force a state change of the
diluent sensor.
018
Copy Protection Violation
The instrument detected wrong resource files on the Program Disk.
019
System Disk File Corrupt
The instrument detected a failure of the CRC check while loading a file from
disk. This failure indicates that the system disk file is corrupt.
020
Count Period Timeout
During one of the 12 count periods, the system was channelizing data for
longer than one second.
021
Software Timer Error
The software was unable to allocate a software timer.
026
Aspirate Syringe Failure
The aspirate syringe was not sensed at the home sensor when it should
have been.
027
Aspirate Syringe Failure
The aspirate syringe was sensed at the home sensor when it should not
have been.
028
Diluent Syringe Failure
The diluent syringe was not sensed at the home sensor when it should have
been.
029
Diluent Syringe Failure
The diluent syringe was sensed at the home sensor when it should not have
been.
030
Probe Mechanism Failure
The probe was not at the top position when it should have been.