Service manual
PN 4237242C
C.1-1
C
C
MESSAGE/ERROR CODE LISTINGS
C.1 INSTRUMENT NON-FATAL ERROR MESSAGES
Table C.1-1 Non-fatal Error Messages
Displayed Message Description
Hgb Blank Voltage High
An Hgb-blank reading was greater than 4.95 V.
Hgb Blank Voltage Low
An Hgb-blank reading was less than 2.50 V.
Unable to Create INF File
The system attempted to create an INF file but was unable to do so.
Vacuum out of Range
The vacuum for one of the 12, one-second count periods was outside 5.83 to 6.17 in. Hg.