Service manual

PN 4237242C
7.2-1
TROUBLESHOOTING
GENERATING A SERVICE REPORT
7
7.2 GENERATING A SERVICE REPORT
The Service Report provides data about the previous aspirate cycle, whether it was a sample, a
control or a calibrator. This data is not displayed on any customer screen or printout.
To access the Service Report, at the Main Menu select
5 SPECIAL FUNCTIONS tt 4 SUPERVISOR tt 7
[no menu item displays on the screen].
A partial report with Aperture Integrity Monitor (AIM) data is displayed on the screen, and if
the Auto Print function is active, a more comprehensive report is automatically sent to the
Printer.
The Screen Display
The top line of the display has two data entry points for RBC and WBC 26-percentile target
values (Figure 7.2-1). For a description of what this screen displays, see the information
under the next Heading, The Printed Report.
Figure 7.2-1 Sample Screen Display
The Printed Report
If the Service Report function is accessed with Auto Print active, a report is automatically sent
to the printer. An example of the printed report is shown in Figure 7.2-2. The information in
Figure 7.2-2 follows the report, heading by heading, and describes the information contained
in the report.
7242046A
C