Service manual

PN 4237242C
5.2-1
MAINTENANCE PROCEDURES
PMI RECOMMENDATIONS
5
5.2 PMI RECOMMENDATIONS
The MD II is a low maintenance instrument that does not require a PMI. Those service
organizations that will perform PMIs on this instrument can consider the following
components and verifications for their procedure.
Components
Filters
The green fluid barriers and blue particle filters should be replaced on every PMI. The
effectiveness of these filters is determined by environmental and instrument conditions more
than by instrument cycle count.
Aged peristaltic pump tubing or contaminated diluent determine replacement of the blue
particle filter. Diluter problems, especially those associated with the VIC, determine
replacement of the green fluid barriers.
Peristaltic Pump Tubing
The peristaltic pump tubing should be changed every 18,000 instrument cycles. Find out the
customer's usage rate to decide whether this tubing should be changed every PMI or not.
Polyurethane Tubing
Polyurethane tubing should be changed when it loses its elasticity. When the tubing becomes
stiff and discolored it has lost its elasticity.
Syringes
The diluent and aspirate syringes should be changed every 35,000 instrument cycles or when
by visual inspection you can see that it needs changing. Even though discoloration caused by
rusting of the shaft may occur before 35,000 instrument cycles, performance should not be
affected.
Cleaning
1. Remove the dust from the upper chassis using a vacuum or brush.
2. Remove the salt deposits from the bath area and shield.
3. Inspect the tubing and fitting connections. Remove any salt deposits and repair any leaks
you find.
4. Bleach the baths and apertures.
C