Service manual
PN 4237242C
4.18-1
SERVICE AND REPAIR PROCEDURES
PROBE MOTOR
4
4.18 PROBE MOTOR
Tools/Supplies Needed
B #1 Phillips-head screwdriver
Removal
1. Turn OFF the instrument’s power.
2. Open the lower chassis as directed in Heading 4.4, to expose the Probe/Wipe Traverse
Assembly.
3. Disconnect the probe motor and the flex cable. The probe motor is connected to J2 on
the Probe/Opto Sensor card (Figure 4.17-1).
4. Make sure the probe is in a vertical position. Move the traverse housing to the right end
of its travel. This aligns it with two probe motor access holes in the traverse mounting
bracket (Figure 4.19-1).
5. Remove the probe motor. It is fastened to the traverse housing with two Phillips-head
screws.
Installation
1. Install the replacement motor, making sure that the motor gear properly meshes with the
traverse rack.
2. Fasten the motor using the two Phillips-head screws. There are screwdriver access holes
at the right end of the traverse mounting bracket (Figure 4.19-1).
3. Connect the motor to J2 on the Probe/Opto Sensor card and reconnect the flex cable.
4. Close the upper chassis and turn ON the instrument’s power.
Verification
Perform an SVP (Heading 5.1).
C