Service manual
PN 4237242C
4.17-3
SERVICE AND REPAIR PROCEDURES
PROBE AND PROBE WIPE
4
5. Install the aspirate tubing onto the probe wipe housing (Figure 4.17-1) if it was
removed. The outside large bore tubing of the three-tube ribbon attaches to the top
fitting while the middle tubing attaches to the bottom. The top of the probe wipe
housing is the end with the groove for the wire retaining clip.
6. Thread the probe wipe housing over the aspirate probe and into the traverse housing,
then install the wire retaining clip to hold it in place (Figure 4.17-1).
7. Close the lower chassis.
8. Turn ON the instrument’s power.
Verification
1. Run samples and ensure that:
r The aspirate probe moves freely and remains in place.
r There are no fluid leaks from the probe wipe and aspirate components.
2. Perform an SVP (Heading 5.1). Control recovery verifies that you have no leaks in the
aspirate tubing or its seal to the aspirate probe.