Service manual

PN 4237242C
4.10-1
SERVICE AND REPAIR PROCEDURES
MEMBRANE KEYPAD AND DISPLAY
4
4.10 MEMBRANE KEYPAD AND DISPLAY
Tools/Supplies Needed
B #2 Phillips-head screwdriver
B 1/4 in. nut driver
Removal
1. Turn OFF the instrument’s power.
2. Remove the top cover as directed under Heading 4.3.
3. Disconnect the following cables (Figure 4.10-1):
a. The display cable (ribbon) from J4 on the URA card.
b. The keypad cable (flex) from J5 on the URA card.
c. The keypad flex cable shield ground from its connection post on the chassis.
4. Remove the eight, #2 Phillips-head screws that fasten the keypad/display assembly to the
chassis (Figure 4.10-1).
Figure 4.10-1 Membrane Keypad and Display
5. Remove the keypad/display assembly from the chassis, being careful not to snag one of
the connectors as you pull the cables through the chassis opening (Figure 4.10-1).
URA
CARD
PHILLIPS-
HEAD
SCREW
(8)
KEYPAD
KEYPAD
CABLE
DISPLAY
CABLE
7242038A
DISPLAY
FLEX CABLE
SHIELD
GROUND
C