English FOR ULTRASONIC NEBULIZER MODEL CUN60 Español Português Deutch INSTRUCTION MANUAL Français - Eng 1 -
FEATURES ★ Handy design - with built-in detachable battery pack; compact and lightweight. ★ Easy to use - uses water as ultrasonic conducting medium, no gel required. ★ Silent operation - thanks to the ultrasonic vibration technology. ★ 3-speed nebulization rate control - provides optimal treat-ment for all kinds of medication with different density. ★ Built-in timer - automatically shuts off after 10-minute use.
PREPARATION ★ Use the equipment only for its intended use as de- tion will be very weak. OPERATION ★ Close supervision is required when this equipment is used by children or invalids. ★ Stop using the equipment immediately if any abnormality occurs. ★ Keep your eyes away from the output of medication mist. ★ Do not use the equipment where flammable gas is being used. ★ Never operate the equipment with empty water tank or medication cup. ★ Do not use solutions other than water for the tank.
★ Never open the clear cover during operation. It may cause a severe injury. ★ Do not use the equipment for more than 30 minutes. Do not reuse within 30 minutes. ★ Some medications may not have good atomization. Check with the physician or manufacturer of medication. Manufacturer: Name: Address: CLEANING ★ Never immerse the equipment in water. It may damage the unit. ★ Clean all necessary parts after each use, as described on page 8.
English PARTS DESCRIPTION Filter cover Air filter Top cover Low-battery indication lamp Main body Nebulization rate Indication lamps Français ON/OFF button Nebulization rate control button Medication cup Fan Rechargeable Battery pack - Eng 5 - Español Português Deutch Water reservoir Mouth nozzle
OPERATION ★ To use the equipment for the first time, or after stor- 2. FILL MEDICATION CUP ing it for an extended period, be sure to clean and sterilize all necessary parts as described on page 8. ★ Always unplug this product and turn off the switch at bottom before filling and after use. a. Replace medication cup on water reservoir. b. Fill medication directly into medication cup. c. Replace top cover. Caution: 1. Do not fill more than 8 ml, otherwise the nebulization rate will be very weak. 2.
5. ADJUST NEBULIZATION RATE & BEGIN TREATMENT a. Press the “nebulization rate control button” to adjust the nebulization rate at high, medium or low speed, which will be illuminated by corresponding indication lamp. b. Begin treatment according to your physician's instructions. Caution: Stop operation immediately if any abnormality is noted. - Eng 7 - Français a. Press the On/Off button to stop operation. b. Turn the switch at bottom to “OFF/CHARGE” position. c.
CLEANING ★ Be sure to clean and sterilize all necessary parts after each use. ★ Always unplug this product before cleaning. ★ Never immerse the equipment in water. It may damage the unit. 1. DISPOSE EXCESS MEDICATION AND WATER 2. CLEANING a. Nose piece, mouth nozzle, extension tube, mask, top cover, medication cup, air filter & filter cover: Clean with mild detergent or commercially available disinfectant. Rinse all parts thoroughly and let air dry. b.
Specifications subject to change without notice. - Eng 9 - Français Auto shut-off timer: Dimensions (L × W × H): Weight: Accessories: 220-240VAC 50/60Hz Approx. 10W 2.5MHz 3-speed adjustable, 0.2 - 0.7 ml/min. depending on medication density 5 µm MMAD Approx. 8 ml max. –10 ~ 50°C; 45 ~ 85% Type B apparatus; Class II insulation Approx. 10 min. 88 × 53 × 143 mm Approx.
TROUBLESHOOTING ★ Should any problem persist after you have made the following checks, return the product to a nearest service center for examination and repair. Symptom Cause and/or solution • Check that AC adapter is plugged in firmly. • Check that switch at bottom is turned ON. Unit does not operate when ON/OFF button is depressed. • Check that water reservoir is filled to the correct level. • If using with battery only, check that battery pack is firmly attached to the unit.
1. CHARGING THE BATTERY a. Install the battery pack by sliding it into the right-bottom of nebulizer. b. Make sure the switch at bottom is turned to “OFF/ CHARGE” position. c. Connect adapter’s plug to nebulizer’s power receptacle. d. Plug the adapter into an AC outlet. e. The required charging time is 24 hours for the first time of use, and 4 to 12 hours thereafter depending on duration of usage. It is recommended that the battery is charged after each use, and let charged overnight.
2. USING THE BATTERY a. Turn the switch at bottom of nebulizer to “ON” position. The LED indicator will be lit. b. Start using the nebulizer as described on page 6. c. After operation is finished, turn the switch at bottom to “OFF/CHARGE” position to save battery power. d. To remove battery pack, press the grooved area and release it from nebulizer by sliding down.
Ni-MH Battery voltage: 12VDC Charging current: 40 ~ 100 mA Charging time: Minimum 24 hours first time; 4 ~ 12 hours thereafter depending on duration of usage Capacity: 480 mAH/12VDC (approx. 30-minute continuous use) 0 ~ 45°C (32 ~ 112°F) Storage temperature: –18 ~ 45°C (0 ~ 112°F) Dimensions (L × W × H): 47 × 28 × 84 mm Weight: Approx. 125 g Français Operating temperature: Specifications subject to change without notice.
●CITIZEN is a registered trademark of Citizen Watch Co., Japan. Consumer Electronics Division CBM Bldg. 5F, 5-68-10, Nakano, Nakano-ku, Tokyo 164-0001, Japan Phone: (03)5345-7430 Fax: (03)5345-7431 E-mail: info-con@jcbm.co.