English FOR DIGITAL BLOOD PRESSURE MONITOR Español INSTRUCTION MANUAL Português MODEL CH-432B Italiano CH -4 SY ST OL IC TO LI 32 B m Hg Cm PULS E /m - Eng 1 - ER RT Français STA POW in Deutch DI AS m m Hg
CLEANING METHOD OF CUFF After cleaning the cuff with a neutral detergent, lightly rinse it with water and let it dry in the air. Hard rubbing or wash- ing in a washing machine will cause air leakage. Also, be careful not to get water into the air hose. General Remarks on blood pressure measurement • This Product should not be used to monitor cardiac arrhythmia. • If the cuff is to be deflated quickly, depress “POWER” switch. • The attached cuff is suited for an arm circumstance of 22cm to 32cm.
• The memory is erased if the batteries are removed for replacement, etc. and a record should be kept of any necessary data. Precautions for maintenance - Eng 3 - Français • Do NOT store the blood pressure monitor where it will be exposed to direct sunlight, high temperature (over 60˚C), low temperature (below –20˚C), high relative humidity (over 95%) or in a dusty place. • Do NOT expose the blood pressure monitor to under force or vibration, or drop it.
• When the monitor is very dirty, wipe it clean with a cloth moistened with Sterilizing alcohol or a neutral detergent, and then wipe it with a dry cloth. Do NOT use thinner or benzene. Nature and frequency of maintenance: This products is designed for use over an extended period of time; however, it is generally recommended that it be inspected every two years to ensure proper function and performance.
Systolic blood pressure display section Diastolic blood pressure/ Pulse display section SY 32 B TO LIC mm Hg mm Hg in E /m PULS BT RT ER Cuff tube Italiano STA POW Português CH -4 DI AS IC STOL Español Tube socket English IDENTIFICATION OF PARTS Air connector Power/Start switch Deutch Battry cover Cuff Français - Eng 5 -
MEASUREMENTS 1. Attaching of Arm Cuff and Posture During Measurement Wind the arm cuff around the left upper arm tightly so that bottom edge of the cuff is positioned about 1 – 2cm above the elbow joint. • When you will wind the cuff, please wind the cuff next to the skin of possible. (If measurement is made by rolling the shirts or jacket up over the arm, this will constrict the upper arm and prevent accurate measurement.) Unfasten and open the cuff as shown in the figure.
Español Press the “POWER / START” switch. DIASTOLIC mmHg Italiano SYSTOLIC mmHg Português Rest the elbow on a table so that the section of the upper arm is at the height of the heart. Relax the lower arm across the table and remain still, since movement - artifacts may falsify the measuring result. Make measurements in either a seated or reclining posture with the palm turned upward. English 2.
SYSTOLIC mmHg When the “POWER / START” switch pressing, the pressure level for measurement will change at optional level over 160 mmHg. But if the “POWER / START” switch is released before 160 mmHg, the pressure is heightened to 160mmHg. DIASTOLIC mmHg PULSE /min SYSTOLIC mmHg SYSTOLIC mmHg DIASTOLIC mmHg DIASTOLIC mmHg PULSE /min When the pressure lowers by 15mmHg after pressurization stops, the mark of “Q” is displayed.
SYSTOLIC mmHg DIASTOLIC mmHg PULSE /min SYSTOLIC mmHg DIASTOLIC mmHg Manual Pressure Setting - Eng 9 - Français When a person with maximum blood pressure of 170 mmHg or higher uses the unit for the first time, or after changing batteries, measure blood pressure using the manual setting. If the automatic pressure setting is used, it may not work. Deutch The measured data are registered automatically, as long as they are not error.
3. Turn Power OFF Battery cover Press the “POWER / START” switch to turn off. Remove the battery cover at mainframe and replace four type “AA” batteries. 4. Subsequent Measurement Batteries are almost empty. When a measurement is need again, 10 minutes, interval is necessary between measurements. Batteries are normal. 5. Automatic Power OFF Function If the unit is left without any operation after measurement, an “Automatic Power-OFF Function” runs to turn off the power after approximately 3 minutes.
DISPLAY MARK DIASTOLIC mmHg SYSTOLIC mmHg Displayed when the pulse rate could not be measured. Take off the cuff and wind it again, and carry out measurement once more. Do not move the arm during measurement. Displayed when the batteries are almost empty. Replace four batteries. The unit has a trouble. Take it to the store where it was purchased for testing.
SPECIFICATIONS Model No.
●CITIZEN is a registered trademark of Citizen Watch Co., Japan. Consumer Electronics Division CBM Bldg. 5F, 5-68-10, Nakano, Nakano-ku, Tokyo 164-0001, Japan Phone: (03)5345-7430 Fax: (03)5345-7431 E-mail: info-con@jcbm.co.