Contents 1. SETTING UP THE BALANCE.............................................................................. 3 2. APPLICATION...................................................................................................... 4 3. CONDITIONS OF APPROPIATE USING ............................................................. 5 4. KEYBOARD ......................................................................................................... 6 5. START AND SERVICE........................................
12. REPORTS FROM SCALE ................................................................................ 33 12.1. Current control of calibration fault ............................................................. 33 12.2. Report from calibration of scale................................................................. 34 13. WEIGHING LOADS UNDER SCALE ............................................................... 35 14. LIST OF ORDERS COMPUTER - SCALE .......................................................
1. SETTING UP THE BALANCE Cut safety tape off. Take balance off the box. Take all necessary to correct performance of balance elements off the box. After replace balance in place of use put scale and rest of elements on. Assembly of elements for scale with precision 10mg: - - Put pan scale (2) on gum mandrels Put cover on scale (1) using pins in back part of balance cover.
2. APPLICATION Electronic scale CY - CG is appropriate for precise measurements mass in laboratory conditions with possibility of setting to zero in all measure range. Balance calibration is done by CAL on balance keyboard. Functions accessible to user are divided into 4 parts: ! Functions for weighing units (press F key) - Weighing in grams, - Result with 3 places after dot, - Weighing in carats [ct], - Weighing in units from beyond SI: [GN], [mom], [oz], [dwt].
3. CONDITIONS OF APPROPIATE USING • scales should be used and stored in rooms without vibrations and shakes, without drafts and dust o o • temperature of air should be +15 C ÷ +35 C • during using balances changes of temperature shouldn’t be higher than 0,5oC per one hour • scales should be replaced on wall console or stable table, far from sources of heating • strong magnes is installed inside of balance – so it’s important to be carefull during weighing magnetic materials.
4. KEYBOARD Key ON/OFF is to turn display off. Others subassemblies are still suppplying and scale is in stage of waiting for further work. If user press this once again weighing stage with measure units are displayed. PRINT key is to send current state of display to external mechanism by connector RS 232 CAL key is to calibrate. It’s manual calibration. After press if scale starts calibration. There shouldn’t be any load on scale.
5. START AND SERVICE Before supply is turn on glass of weighing chamber should be draw Balance should be leveled. To level scale screw legs in order air bubble was in centre position. TIME OF HEATING It’s neccesary to wait untill scale gets heat stabilization. It’s time of own heating. User has to wait 15 min. in case of analitycal scales. Given values refers to scales that were in temperature of conditions of work before they are connected to suppling net. If scales were in much lower temperature (f.eg.
6. WEIGHING Before weighing or if the conditions of work changes (f.eg. if temperature of surrounding is higher then 3oC) scale should be calibrated according to way in point 7.
7. CALIBRATION To assure very precise weighing introduce correction factor to balance memory – it’s calibration.
7.1. Balance with external calibration Pressing key CAL causes calibration of balance - on display appears: (balance starts autotest for calibration –during test you cannot load the scale), and then order (it is command to put calibration mass on scale) calibration mass is given in grams. If you the mass on scale on display appears calibration and then take calibration mass off scale. After take calibration weight off scale balance shows zero.
8. SCHEDULE OF ENTERS IN SETTING FUNCTIONS - Counting pieces Control of deviations % according to standard mass Thickness of liquids Thickness of solids Press TARE key (scale shouldn’t display STAB). If horizontal lines appears on display press key F After pressing key F name of first approachable function appears. Every next pressing key F makes names of following approachable functions appear on display. Way of setting functions is described in further part of instruction.
8.1. Counting pieces Softwares enables to amount small elements with the same mass To use this function: • put package on scale and press TARE its mass to balance memory • press key TARE. When horizontal lines appear on display press key F • Press F key – display appear • Press PRINT key to activate function (you can select from 10, 20, 50pcs or free definable) • • Press F key – choose value If you want to start the counting pieces function, please choose one of the values, eg.
• • After displaying of LoAd information, please put on the pan of the balance as many pieces as have been chosen (20pcs).
• set up batch quantity by mens of keys ON/OFF – choose digit F – choose value • confirm by pressing key PRINT LOAD appears on display • after making sure that batch with declared quantity is put on scale confirm by key PRINT once again - balance displays Process, and then amount of elements and PCS.
Return to weighing mode • press key F many times untill - PIECES appears on display • press TARE key Balance return to weighing mode and diode with g mark starts flashing Attention: 1. 2. 3. If you choose batch quantity remember that precision of counting depends on metrological parameters of balance and on repeatability of mass singular counted elements.
8.2. Control of deviations % according to standard mass Software enables control of deviations (in %) mass of weighed loads from standard mass. Standard mass can be defined by weighing (function PERC A) or written to balance memory by user(function PERC B) 8.2.1. Standard mass defined by its weighing Procedure: Press key TARE.
8.2.2. Standard mass introduced to balance memory by user Procedure Press key TARE.
8.3. Definition thickness of solids and liquids There is equipment to defining thickness of solids and liquids in additional equipment. For user request Citizen service can render special software to define thickness by scale accessible. 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. beaker support frame of scale plunger beaker thermometer clench thermometer plunger string plunger hanger upper scale scale string down pan additional loads for leveling initial load 8.3.1.
9. Adjusting the balance to working conditions - Turn off/on system of autozero - Speed of stabilization - Average result of weighing - Adjusting balance for dosage procedure - Select default unit of weighing Press TARE key (scale shouldn’t display STAB). If horizontal lines appears on display press key CAL After pressing key F name of first approachable function appears. Every next pressing key F makes names of following approachable functions appear on display.
9.1. Turn on/off autozero system Press key TARE. When horizontal lines appears on display press key CAL. Then press key F many times untill order AUTO appears on display. Press key PRINT • Current value for autozero function appears on display. • Press key F many times and sett value on 0 or 1 including following dependences: • AUTO 0 - autozero is turn ON AUTO 1 - autozero is turn OFF Press key PRINT to confirm setting.
9.2. Choice of criteria of stable result depending on conditions of using Press key TARE. When horizontal lines appears on display press key CAL. Then press F key many times until order Con appears on display • Press PRINT key. Next to Con settled value of parameter with flashing number is displayed. Flashing number means stage of average result • By pressing key F change value of flashing number choosing one of the stages of averaging.
9.3. Average result of weighing In various applications result of measurement is presented as average value of several measurements especially if the conditions are not stable (vibrations, twitches). In these cases if you use higher stage of the digital filter the measurement is done with the established precision. Press the TARE button, when horizontal lines appear on the display press the CAL button. • • • The balance shows first function in the menu.
9.4. Adjusting scale for dosage procedures During dosage substances essential matter is fast reaction of the balance on mass increment. During this process digital filters must be turn off because they make reaction time longer although they help if any disturbances occur. Procedure Press the TARE button, when horizontal lines appear on the display press the CAL button • • • • The balance shows first function in the menu. Press the F key repeatedly until command FIL appears on the display.
9.5. Select default unit of weighing Press the TARE button, when horizontal lines appear on the display press the CAL button The balance shows first function in the menu. Press the F key repeatedly until command St_uni appears on the display. Press PRINT button many times. The display will show units. Press TARE key to confirm selected units.
10. FUNCTIONS FOR USING RS 232 - Choose speed of transmission - Define sent data as: stable / not stable - Continous work - Setting data as additional element of parameter in printouts Press TARE key. When horizontal lines appear on display press key PRINT. After pressing PRINT name of first approachable function appears. Every pressing of key F makes makes names of following approachable functions appear on the display. Way of setting functions is described in further part of instruction.
10.1. Choice of speed of transmission Press key TARE. When horizontal lines appear on display press key PRINT. Then press key F untill order bod appears on display. Press key PRINT. • Press PRINT key (diplay shows value for function BOD) • press F key many times untill number of speed of transmission appears as below: • bod 1 - 300 bit / s, bod 2 - 600 bit / s, bod 3 - 1200 bit / s, bod 4 - 2400 bit / s, bod 5 - 4800 bit / s, bod 6 - 9600 bit / s, confirm by pressing PRINT.
10.2. Definite sent data as stable/not stable Press key TARE. When horizontal lines appear on display press key PRINT • Then press key F many times until order StAb appears on display. • by means of PRINT key start programming functions • by key F choose one of two values STAB STAB • 0 – sending stable result of weighing 1 - sending temporary result of weighing confirm by pressing PRINT.
10.3. Continuous work (RS 232) Press key ZERO/TARE. When horizontal lines appear on display press key PRINT. • Then press key F many times until order cont appears on display.
10.4. Set-up date as basic parameter of printout Signal (equvivalent for state of display) is sent to printer by scale has form: + 123.456 g < LF> ( 16 signs in general) There is possibility of sending data in increased form f.eg.: DATE 1998.09.21 TIME 09:51 + 100.001 g Printing results of measurements in increased form is possible only with printer CITIZEN CPR02 Press key TARE. When horizontal lines appear on display press key PRINT.
11. COOPERATION WITH PRINTER OR COMPUTER Every pressing of key makes sending signal of current state of display with measure units to computer or sprinter. Scale has factory settled speed of transmission 4800 bit/s. If external mechanism (printer, computer) need to have settled different speed of transmission change factory setting about speed in menu (parameter bod 11.1.
11.2. Printing data with date and time Every printout of weighing can be printed with date and time of measurement. It’s possible if scale is connected to printer CITIZEN CPR02 After connecting scale and printer sett value of parameter parameter DATE as: DATE 1 (see Setting date as additional parameter for printout) 11.3. Cooperation with statistic printer After connecting scale to statistic printer statistic of measurement is possible to do.
11.4.1. Format of sent data after pressing the PRINT button 11.4.2.
Give any result of weighing 12. REPORTS FROM SCALE Scales with internal calibration can control parameters remembered in memory very fastly. Their function can be increased. Their service functions can be done easily. 12.1. Current control of calibration fault Described function is usefull if during long measurement user wants to be sure that results don’t have bigger errors than terror shown on display. When scale is in autozero zone (ZERO) press keys TARE and F at the same time.
12.2. Report from calibration of scale • • Connect printer with real time clock to scale (CITIZEN CPR02) Press keys TARE and CAL at the same time • Procedure of the reports ! Balance starts autotest for reports. ! Display shows calibration mass ! Put the load on the pan. ! Balance is in auto adjusting mode ! When adjusting is finished, remove load from the pan ! Display show communicate: report ! At the moment display show calibration mass ! Put the load on the pan.
13. WEIGHING LOADS UNDER SCALE In standard analitycal scales can weigh load on suspension. To use this function: • Remove plastic plug in basic of scale • There is suspension in the basic of scale. It’s installed for good. • Install hook to hang load (hook isn’t standard equipment for scale), weigh load on hook. Attention: • Suspension cannot be turned around, move or manipulate. Mechanism of scale can be damaged this way.
14.
15. ORDERS ON DIGITAL DISPLAY Err- 5 FuLL – 1 FuLL- 2 nuLL tESt CAL STAB DATE CONT AUTO - Co - Li tE_L 00 H2O C2H5OH AnotHEr VOLUME LoAD A LoAD L Err-3 CAL Err bod AC_t AuE FIL St_uni - signalling terror of temperature conveter, to start scale once again turn on and off power supply - overcrossed range of conveter A/D - overcrosed range of weighing - too small load of scale (np.
16. TECHNICAL PARAMETERS Type CY 110 CY 220 CY 360 CY 510 CY 720 Max. capacity 110g 220g 360g 510g 720g Min. capacity 20mg 40mg Readability 1mg 2mg Tare range -110g -220g -360g -510g -720 + 15oC do +35oC Operat. Temp 230V 50Hz AC / 10,5 V AC Power 128 x 128mm Pan size Type CG 1202 CG 2202 CG 3102 CG 4102 Max. capacity 1200g 2200g 3100g 4100g -3100g -4100g Min. capacity 500mg Readability 10mg Tare range Operat.
17. ADDITIONAL EQUIPMENT 17.1. Vibration damping table Thank’s to this stable base all vibrations are eliminated.Inside of table marble plate is installed which makes foundation for balance. 17.2. Rack for weighing load under balance It is used when weighing under balance is neccesary. It is usefull for magnetic load or when thickness of products is defined. Racks are used also during tests for absorptiveness of materials f.eg. absorptiveness of foamed polystyrene. Structre of rack is made of steel.
!!