Specifications
Etigam BRIGHT-CHEQ Biological Indicator Monitoring System
BRIGHT-CHEQ Incubator Operation Manual
6
UNPACKING AND INSPECTION
1. Remove all contents from the box and carefully inspect all components for any damage
caused in transit. If any damage is observed, notify your Etigam representative
immediately.
2. Verify that the following components have been included in the kit:
• BRIGHT-CHEQincubator
• BRIGHT-CHEQ12Vpowerpack–inputvoltage110or220
• NISTThermometer
• Printerkitincludingprinter,manual,7Vpowerpack(inputvoltage110or220),plug
adapter for power pack (optional), one roll of paper, and one printer ribbon
• Serialprintercable
• Stylus
• BRIGHT-CHEQrecordbook
• BRIGHT-CHEQ Incubator Manual and Etigam BRIGHT-CHEQ Biological Indicator
Monitoring System Product Overview
OPERATIONAL PRECAUTIONS
!
Use Only as Directed
• Indooruseonly.
• Temperatures5°Cto37°C.
• Altitude<2000M.
• PollutionDegree2.
• InstallationCategoryII.
• Maximum relative humidity 80% for temperatures up to 31°C
decreasinglinearlyto50%relativehumidityat37°C.
• Mainpowersupplyvoltageuctuationsnottoexceed±10%ofthe
nominalvoltage.
• Do not place incubator in drafty areas, near open windows, air
conditioning, heating vents, or heat-generating equipment since
uctuatingtemperaturesmayaecttheincubatortemperature.
• Incubator is designed for use with Etigam BRIGHT-CHEQ self-
contained biological indicators containing G. stearothermophilus.
Erroneous results may occur when other biological indicators are
used.
INSTALLATION