Specifications

Etigam BRIGHT-CHEQ Biological Indicator Monitoring System
BRIGHT-CHEQ Incubator Operation Manual
6
UNPACKING AND INSPECTION
1. Remove all contents from the box and carefully inspect all components for any damage
caused in transit. If any damage is observed, notify your Etigam representative
immediately.
2. Verify that the following components have been included in the kit:
• BRIGHT-CHEQincubator
• BRIGHT-CHEQ12Vpowerpack–inputvoltage110or220
• NISTThermometer
• Printerkitincludingprinter,manual,7Vpowerpack(inputvoltage110or220),plug
adapter for power pack (optional), one roll of paper, and one printer ribbon
• Serialprintercable
• Stylus
• BRIGHT-CHEQrecordbook
• BRIGHT-CHEQ Incubator Manual and Etigam BRIGHT-CHEQ Biological Indicator
Monitoring System Product Overview
OPERATIONAL PRECAUTIONS
!
Use Only as Directed
• Indooruseonly.
• Temperatures5°Cto37°C.
• Altitude<2000M.
• PollutionDegree2.
• InstallationCategoryII.
• Maximum relative humidity 80% for temperatures up to 31°C
decreasinglinearlyto50%relativehumidityat37°C.
• Mainpowersupplyvoltageuctuationsnottoexceed±10%ofthe
nominalvoltage.
• Do not place incubator in drafty areas, near open windows, air
conditioning, heating vents, or heat-generating equipment since
uctuatingtemperaturesmayaecttheincubatortemperature.
• Incubator is designed for use with Etigam BRIGHT-CHEQ self-
contained biological indicators containing G. stearothermophilus. 
Erroneous results may occur when other biological indicators are
used.
INSTALLATION