VIEWING CELL STATUS
Main Menu
Cell Menu
Query Screen
1. From the Main Menu, touch the key
representing the cell to be viewed to enter its
CellMenu.
2. From the CellMenu, touch the QUERY key.
The status of the cell will appear, including
the cell number, the status of the current test
(TESTING | + | -), the start time and date (ST),
the amount of time that the test has been
running (TT), the cycle type and exposure
time, the operator initials (Usr), the Etigam
BRIGHT-CHEQ BI lot number (Lot), and the
sterilizer number (Ster).
3. Touch anywhere on the screen to return to the
CellMenu.
3. Touch the BACK key to return to the Main
Menu.
POWER FAILURE RECOVERY
Power Interrupt Printout
If the BRIGHT-CHEQ incubator loses power, any test in progress
will be saved and upon power return the unit will resume
incubation and monitoring. The unit will automatically
adjust incubation time to make up for the duration of the
power loss. A printout will be generated indicating the time
that the power failure occurred and the time that power was
returned.