Specifications

Etigam BRIGHT-CHEQ Biological Indicator Monitoring System
BRIGHT-CHEQ Incubator Operation Manual
24
Figure 1. BRIGHT-CHEQ incubator
TESTING ETIGAM BRIGHTCHEQ BIOLOGICAL INDICATORS
1. To activate a Etigam BRIGHT-CHEQ BI, place it into
the BRIGHT-CHEQ incubator crushing cavity (Figure
1) and pull it forward until the glass ampoule inside
breaks, releasing the purple growth media.
2. Place one unexposed, activated BI into the positive
control cell labeled C (Figure 1). A yellow BI is the
desired result for the positive control. A positive
control is to be used anytime a load is being veri ed
or at least once per day.
3. Activate all processed Etigam BRIGHT-CHEQ BIs and
place them into the incubation cells numbered 1
through 10 (Figure 1).
When a Etigam BRIGHT-CHEQ BI is placed into a cell, the status LED (Figure 1) in front of
the cell will illuminate amber indicating that the incubator has recognized the BI. The
corresponding key on the LCD MainMenu (Figure 2) will display the word TEST and
the number of whole hours that the test has been in progress (Figure 3). (e.g. if a test
has been in progress for four and a half hours, the display will read 04H).
Figure 2. Main Menu with positive,
negative, and testing samples
Figure 3. Cell key
Table 1. Status LEDs color descriptions
Status LEDs Color Descriptions
O Cell Empty
Amber Test in Progress
Blinking Amber Test in Progress, BRIGHT-CHEQ removed (max. 10 seconds)
Red Test Positive
Blinking Red Test Positive, BRIGHT-CHEQ removed (max. 10 seconds)
Green Test Negative
Blinking Green Test Negative, BRIGHT-CHEQ removed (max. 10 seconds)
All Colors Cycling Invalid Test, BRIGHT-CHEQ must be removed
If the incubator does not recognize a BRIGHT-CHEQ BI
that has been placed into an incubation cell, give the BI
a quarter turn or try the BI in a di erent cell. Use only
Etigam BRIGHT-CHEQ biological indicators.
Refer to Appendix A. Troubleshooting.
OPERATION