DUPLICATING CELL CONFIGURATION
Main Menu
Cell Menu
Setup Cell Menu
Setup Cell Menu Page 2
1. From the MainMenu, touch the key representing
the cell that conguration settings will be copied
from to enter its CellMenu.
2. From the CellMenu, touch the SETUP key to enter
the SetupCellMenu.
3. Touch the MORE key to advance to Setup Cell
MenuPage2.
4. To select the destination cell(s) to which the
conguration settings will be copied, touch
the DEST toggle key until the desired selection
appears.
The DEST key toggles between incubation
cells represented by numbers 1 through 10, the
positive control cell represented by the word
CTRL, and all eleven cells represented by the
word ALL.
Destination Cell Options
ALL 1 2 3 4 5 6 7 8 9 10 CTRL
5. Touch the COPY key to copy the information to the
selected cell(s).
6. Touch the BACK key to return to the CellMenu.
7. Touch the BACK key to return to the MainMenu.
8. To verify that conguration settings were copied
successfully, query each cell that parameters were
copied to. Refer to the Viewing Cell Status and
Printing procedures in the Operation section of this
manual.
9. Repeat this procedure for other cells as necessary.
Touching the BACK key from either the
SetupCellMenu orSetupCellMenu
Page2 will return to the CellMenu.
CONFIGURATION - CELL SETTINGS