Specifications

Etigam BRIGHT-CHEQ Biological Indicator Monitoring System
BRIGHT-CHEQ Incubator Operation Manual
22
DUPLICATING CELL CONFIGURATION
Main Menu
Cell Menu
Setup Cell Menu
Setup Cell Menu Page 2
1. From the MainMenu, touch the key representing
the cell that conguration settings will be copied
from to enter its CellMenu.
2. From the CellMenu, touch the SETUP key to enter
the SetupCellMenu.
3. Touch the MORE key to advance to Setup Cell
MenuPage2.
4. To select the destination cell(s) to which the
conguration settings will be copied, touch
the DEST toggle key until the desired selection
appears.
The DEST key toggles between incubation
cells represented by numbers 1 through 10, the
positive control cell represented by the word
CTRL, and all eleven cells represented by the
word ALL.
Destination Cell Options
ALL 1 2 3 4 5 6 7 8 9 10 CTRL
5. Touch the COPY key to copy the information to the
selected cell(s).
6. Touch the BACK key to return to the CellMenu.
7. Touch the BACK key to return to the MainMenu.
8. To verify that conguration settings were copied
successfully, query each cell that parameters were
copied to. Refer to the Viewing Cell Status and
Printing procedures in the Operation section of this
manual.
9. Repeat this procedure for other cells as necessary.
Touching the BACK key from either the
SetupCellMenu orSetupCellMenu
Page2 will return to the CellMenu.
CONFIGURATION - CELL SETTINGS