Specifications
BRIGHT-CHEQ Incubator Operation Manual
Etigam BRIGHT-CHEQ Biological Indicator Monitoring System
21
SETTING THE STERILIZATION CYCLE TYPE, STERILIZATION
TEMPERATURE, AND STERILIZATION EXPOSURE TIME
Cell Menu
Setup Cell Menu
Cycle Setup Menu
1. From the Main Menu, touch the key representing
the desired cell to enter its CellMenu.
2. From the CellMenu, touch the SETUP key to enter
the SetupCellMenu.
3. From the SetupCellMenu, touch the CYCLE key to
enter the CycleSetupMenu.
4. To set the cycle type, touch the TYPE toggle key
until the desired setting appears.
Cycle Type Options
Prevac
Flash
Express
Gravity
5. To set the cycle exposure time, touch the TIME key.
Using the keypad that appears on-screen, enter the
cycle exposure time in minutes. Use the ARROW
keys to move the cursor position, if necessary. To
save and return to the Cycle Setup Menu, touch
the ENTER key. To return to the CycleSetupMenu
without saving, touch the CANCL key.
6. To set the cycle exposure temperature, touch the
TEMP toggle key until the desired setting appears.
Cycle Temp. (°C) Options
121° 132°
134° 135°
7. Touch the BACK key to return to the Setup Cell
Menu.
8. If cell conguration is complete, touch the BACK
key to return to the CellMenu, then touch the BACK
key again to return to the MainMenu.
CONFIGURATION - CELL SETTINGS