Instruction manual
Chapter 2 Before Measurement (Meter SET)
32 HORIBA
2.14 Analog Output Adjustment
How to Analog Output Adj.
1.
Tap on the right of the Analog output adj.
item.
The Analog output adj. screen is displayed.
Tap to return to the previous screen.
1. Tap on the right of the analog output item.
The Output value adjustment screen is dis-
played.
2. Tap or to adjust the analog output volt-
age.
3. Tap .
To cancel the operation, tap to return to the
previous screen.
Voltage output can be acquired from the analog
output connector located at the instrument side.
Connect the instrument with a digital multimeter,
digital recorder, pen recorder or the like using an
designated cable (analog output cable: Parts
No.3014030152), and check and adjust the
analog output value of the instrument.