Specifications
5-28
Cisco ONS 15216 EDFA Operation Guide
78-12818-01
Chapter 5 Command Line Reference
Command Summary
SETGT Sets laser diode
pump control modes
to constant gain
This command sets the laser
pump control modes to constant
gain with possible temperature
compensation. When switching
laser diode pump control modes
to constant gain, the loss of
output alarm, LPOUT, is
disabled and the GAIN alarm is
enabled.
SETI 1|2 <value,
decimal (0-300)
mA)>
Sets specified laser
diode pump control
to constant current
mode
This command sets Laser 1 or 2
to Constant Pump Laser Current
at set <value> in mA.
SETOUT <value,
decimal (0-65 mW)>
Sets EDFA in
constant output
power mode
Constant EDFA output power
mode has the first-stage pump
laser configured for 75mW of
constant pump power and the
second-stage laser controlling
the overall EDFA output power.
The SETOUT command is
used to set the control point for
the second-stage laser. The
LPOUT alarm is automatically
enabled and the GAIN alarm is
automatically disabled when
you successfully issue the
SETOUT command.
A value of zero for the EDFA
output power setpoint puts both
pumps into IDLE mode. In this
case, no control mode line
displays after you issue the
SETOUT command.
SETP 1|2 <value,
decimal (0-100
mW)>
Sets specified laser in
constant pump power
mode
The GAIN alarm is
automatically disabled when
the SETP command is
successfully issued.
A value of zero for the pump
power setpoint places the pump
in IDLE mode. In this case, no
control mode line displays after
the SETP command.
Table 5-17 Command Summary (continued)
Command Syntax Function Description