User Guide

14
Figure 4 - Near Field Target
6mm
1mm
2mm
3mm
4mm
5mm
6mm 6mm 6mm
The depth of the dead zone may be measured as follows:
1. Apply coupling gel to the scanning surface or ll the water trough with tap
water.
2. Position the transducer above the near eld resolution target and per-
pendicular to the wires. (The wires should appear as dots, not lines).
3. Adjust the instrument settings (gain, TGC, output, etc.) to maximize reso-
lution in the near eld. Record these settings for use on subsequent test-
ing.
4. Freeze the image while the near eld targets are clearly displayed.
5. Measure Dead Zone distance using one of these two methods:
Count how many wires of the near eld target you can see.
Subtracting this number from the total number of targets gives
you the dead zone measurement. For instance, if 3 targets are
visible, the dead zone distance = 2 mm (5mm-3mm).
Use the electronic calipers to measure the distance between the
transducer face and the closest wire target to be resolved from
the reverberation. If the rst target to be resolved is at 4 mm,
then the dead zone distance is “something less than 4 mm”.
6. Record this distance and compare with baseline measurements.
The near eld group consists of parallel, 100 micron diameter, nylon, monolament
wires horizontally spaced 6 mm apart from center to center (Figure 4, page 13). Ver-
tical distance from the center of each wire to the top edge of the scanning surface
ranges from 5 mm down to 1 mm in 1 mm increments.
DEAD ZONE ASSESSMENT CONTINUED