User Guide
10
Figure 3 - Combined Axial/Lateral Resolution Target at 10.5 cm depth (top) and a table listing the distances
between them (bottom)
Targets
C1-D1 C2-D2 C3-D3 C4-D4 C5-D5
Axial
Resolution (mm)
1.0 2.0 3.0 4.0 5.0
Targets
C1-C2 C2-C3 C3-C4 C4-C5 C5-C6
Lateral
Resolution (mm)
5.0 4.0 3.0 2.0 1.0
ELEVATIONAL TESTING
A full characterization of system resolution requires measurement of elevational res-
olution, or slice thickness. Slice thickness is typically much coarser than axial and
lateral resolution as most ultrasound transducer arrays are mechanically focused in
the thickness dimensions.
Two methods are described for measuring elevational resolution. The rst method,
rst described by Skolnick
(1)
, uses the vertical wire targets as follows:
1. Apply coupling gel to the scanning surface or ll the water trough with tap
water.
2. Adjust the instrument settings (gain, TGC, output, etc.) as for a “normal” tech-
nique. Record these settings for use on subsequent testing
3. Orient the transducer to image the length of the vertical target wires, taking
care to adjust the tilt so that the wires are lined up in a vertical column.
4. Rotate the transducer 45° so that only a partial length of the wires is now vis-
ible.
5. Freeze the image and measure the length of each wire segment with the elec-
tronic calibers.
6. Record the measurements.
D5
D4
D3
D2
D1
C1
C2C3C4C5C6
Axial Resolution
Lateral Resolution
Edge to
adjacent Edge
3 mm
Edge to
adjacent Edge
2 mm
1.Skolnick, ML. “Estimation of ultrasound beam width in the elevation (section thickness) plane.” Radiology. 1991 Jul;180(1):286-8.