Specifications

AUDIO FUNCTION GENERATOR(S)
OSCILLOSCOPE
DISPLAY
Traces 2
Trace Types Live, captured, accumulated
Markers 2
Marker Functions Time with amplitude, deviation or
% depth
Delta marker (including 1/D t,
e.g. Hz)
DIGITAL MULTIMETER
VERTICAL
3 dB Bandwidth 16 MHz
Frequency Range DC to 4 MHz (40 MS/s
sampling rate)
Input Range 0 to 100 Vpeak Max, Category II
Scales 2 mV to 20 V/division in a 1, 2, 5
sequence (8 [h] x 10 [w] graticule dis-
play)
WAVE SHAPE Sine, Square, Triangle, Ramp, Digital Coded Squelch,
DTMF
FREQUENCY
Range Sine: 20 Hz to 40 kHz (usable 1 Hz to 40 kHz)
Square, Triangle and Ramp: 20 Hz to 4 kHz (usable
1 Hz to 40 kHz)
Resolution 0.1 Hz
Accuracy ±50 ppm, ±10 ppm typical
LEVEL
Range 1 mV to 5V RMS into a 10 kW load
Resolution 0.1 mV
Accuracy ±1% of setting (10 kW load)
Impedance <10 W
Spectral
Purity
<0.5% (1 kHz, 5 Vrms, 80 kHz BW, 10 kW load,
Sine)
<1.0% (Typical, 20 Hz to 20 kHz, 100 mV to 5
Vrms, 80 kHz BW, 10 kW load, Sine)
AUDIO FILTERS (CHARACTERISTIC RESPONSE)
Filter Type Ripple -1 dB -60 dB
NONE No
Filter
300 Hz Low-
Pass
<0.23 dB, above
20 Hz
330 Hz 590 Hz
5 kHz Low-
Pass
<0.02 dB, above
20 Hz
5.5 kHz 6.7 kHz
15 kHz Low-
Pass
<0.01 dB, above
20 Hz
16.1 kHz 17.8 kHz
20 kHz Low-
Pass
<0.01 dB, above
20 Hz
20.4 kHz 21 kHz
0.3 to
3.4 kHz
Band-
Pass
<1.7 dB 320 Hz/
3.8 kHz
60 Hz/
5.2 kHz
0.3 to
5 kHz
Band-
Pass
<1.7 dB 320 Hz/
5.2 kHz
60 Hz/
9.6 kHz
0.3 to
15 kHz
Band-
Pass
<1.7 dB 320 Hz/
16.1 kHz
60 Hz/
19.9 kHz
0.3 to
20 kHz
Band-
Pass
<1.7 dB 200 Hz/
20.4 kHz
60 Hz/
21 kHz
PSOPH
C-MSG
Band-
Pass
Per C-MSG Spec Per C-MSG
Spec
Per C-MSG
Spec
PSOPH
CCITT
Band-
Pass
Per CCITT Spec Per CCITT
Spec
Per CCITT
Spec
300 Hz High-
Pass
<1.7 dB 320 Hz 60 Hz
Accuracy 5% of full scale (DC to 1 MHz)
10% of full scale (1 to 4 MHz)
Resolution Better than 1% of full scale
Coupling DC, AC, GND
HORIZONTAL
Sweep Factors 1 µSec to 1 Sec/division in a 1, 2, 5
sequence
Accuracy >1.5% of full scale
Resolution >1% of full scale
Input Impedance 1 MW, 20 pF
TRIGGER
Trigger Source Trace A, Trace B, EXT, (or Trace C with
no CH1 or CH2 Input)
Trigger Edge Rising/falling
Trigger Mode Auto/normal
Continuous/single shot
External Trigger Level Hi-Z BNC input on the rear panel of the
unit
Adjustable from -5 to +5 V
AC/DC VOLTMETER
Full Scale Range 200 mV, 2 V, 20 V, 200 V, 2000 V,
Auto
(150 VAC RMS, or VDC MAX input,
Category II)
Resolution 3-½ digits (2000 counts)
Accuracy DC ±1% Full Scale ±1 count
AC ±5% Full Scale ±1 count
AC Volts Frequency Range 50 Hz to 20 kHz
AC/DC AMMETER
Full Scale Range 200 mA, 2 A, 20 A, Auto (20 A range
uses optional shunt connected to
Voltmeter)
Maximum Open Circuit
Input Voltage
30 Vrms referenced to common or
earth ground, Category I
Resolution 3-½ digits (2000 counts)
Accuracy ±5% Full Scale ±1 count
AC Volts Frequency Range 50 Hz to 10 kHz
OHMMETER
Full Scale Range 200 ohms, 2 kohms, 20 kohms, 200
kohms, 2 Mohms, 20 Mohms, Auto
Resolution 3 ½ digits (2000 counts)
Accuracy ±5% Full Scale ±1 count
15