Owner`s manual

– 4 –
Capacitance: 4n-40n-400n-4µ-400µF
±(5% + 10), ±(8% + 10) (above 100µF)
(Test voltage: 3V / Test frequency: 4n-40nF 1KHz; 400n-4µF
70Hz; 400µF 7Hz)
Frequency: 4K-40K-400K-4000KHz ±(0.5% + 1)
(Autoranging) (Sensitivity: 0.2Vrms minimum)
Inductance: 4m-40m-400m-40H ±(5% + 10)
4H ±(5% + 20)
(Test frequency: 4m-40mH 1KHz; 400mH-4H 70Hz; 40H 7Hz)
Transistor HFE: 0-999 HFE
Test base current: 10µA / Test voltage: 3.0 ± 0.4V
Diode : Test current: 1.0 ± 0.6mA / Test voltage: <3.5V
Logic: Logic High 2.8V ± 0.8V / Logic Low 0.8V ± 0.5V
Frequency Response: 20MHz
Detectable Pulse Width: 25nS
Indication: 40msec beep at logic low
Overload Protection
Voltage: 500VDC / 350VAC - 400mV ranges
1000VDC / 750 VAC - 4-40-400-1000/750V ranges
Current: 0.5A / 250 V Fuse for 40m-400mA ranges
10A / 600 V Fuse for 10A range (Maximum 60 seconds)
Resistance: 500VDC / 500VAC
Capacitance: 3VDC / 3VAC rms
Frequency: 500VDC / 500VAC rms
Inductance: 3VDC / 3VAC rms
Logic: 500VDC / 500VAC rms
Diode Test: 500VDC / 500VAC rms
MEASUREMENTS
Please take a few minutes to read the instructions and warnings prior to use.
1 DC VOLTAGE MEASUREMENT
!
WARNING: Maximum input is 1000V DC.
1-1 Plug the red test lead into the V jack and the black test lead into the
COM jack.
1-2 Set the DC/AC switch so that DC is displayed on the LCD.
1-3 Set the range switch to the desired V voltage setting. If you are unsure of
the voltage being measured, select the highest range and reduce until a
satisfactory reading is obtained.