High voltage AC and DC circuits are dangerous. Use caution when working
above 60VDC or 30VAC rms as these voltages pose a shock hazard. Always
consider circuits to be energized. Never assume any equipment is de-
energized. Lack of caution or poor safety practices can be very dangerous
and/or lethal.
SPECIAL FUNCTION
MAX - When MAX is pressed, the last reading is held on the display until a
higher reading is recorded. When using the MAX position on the Frequency
range, the autoranging feature is disabled.
SPECIFICATIONS
General
Display: 3 3/4 digit, 17mm (0.67") H LCD
Maximum Display: 3999
Ranging: Manual ranging / Autoranging on Frequency
Polarity: Automatic, minus (-) sign indicates for negative polarity
Measuring Rate: 2.5 times/second
Input Impedance: 10 M (DCV/ACV)
Operation Temp.: 0 to 45°C
Storage Temp.: -15 to 50°C
Fuse: AGX 0.5A/250V & TUE10A/600V
Power Source: 9V Battery, NEDA 1604 or IEC 6F22
Dimensions: 90(W) x 200(H) x 40(D)mm
Weight: 400g (including battery)
Ranges and Accuracy
DC Voltage: 400mV-4-40-400-1000V ±(0.5% + 1)
(Maximum measurement = 1000V)
AC Voltage: 400mV-4-40-400V ±(1.5% + 4), 750V ±(2% + 4)
(Maximum measurement = 750V RMS)
(50 Hz - 500Hz)
DC Current: 40mA-400mA ±(1.5% + 1), 10A ±(3% + 1)
(Maximum voltage drop <325mV on all scales except 10A
<700mV)
AC Current: 40mA-400mA ±(2% + 4), 10A ±(3.5% + 4)
(Maximum voltage drop <325mV on all scales except 10A
<700mV)
Resistance: 400 ±(1% + 3), 4K-40K-400K-4M ±(0.8% + 2)
40M ±(3% + 4), 4000M ±((5% + 10) +10 )
(Test voltages:400-4000M 3.0V; 4K-40K-400K-4M-40M 0.6V)