User`s manual

NexSens Technology, Inc.
iSIC User‟s Manual 95
Logging:
The Logging tab allows you to view the memory capacity, memory page structure, amount of
memory used, data record size, and display an estimate of the amount of time left before the
memory is full.
Erase logging memory: Usually does not need to be done unless you do not want old data
uploaded into a new iChart database. By default the iSIC will store the last 150,000 readings.
Additionally, if the iSIC is configured for “Stop” log memory mode, the memory must be erased
after each upload to avoid memory filling up in the iSIC and preventing new data from being
logged.
Upload Flash: If there are any problems with uploading the data, or it appears that the
memory is corrupted, the entire flash memory of the data logger can be uploaded for analysis.
The file can be uploaded in full, or only part of the flash can be uploaded. This feature is never
required unless directed to do so by NexSens technical support.