iSIC User‟s Manual 85
4.2 Grounding Protection
Electronic circuits are susceptible to voltage surges. The iSIC has on-board transient voltage
suppression devices. Connect the ground lug to a proper electrical ground (i.e. earth ground or a
properly grounded pipe/metal fixture).
For outdoor applications, use the copper wire supplied with the A38 ground kit to connect the
ground lug to the rod‟s ground clamp. The rod should be driven into the ground as deep as
possible and located near the equipment. In addition, make sure the copper wire has no sharp
bends and is as short as possible. Be sure to locate the clamp on the rod before driving it with a
sledgehammer or fence post driver.
A38 Features:
4 ft. copper coated
ground rod
8 ft. of #8 solid
copper wire
Rod/Wire clamp
Comes assembled
Installation:
Bury copper ground
rod into the earth.
Attach copper wire to
ground lug directly
onto the iSIC or to
the A39 lightning
protector.
For indoor applications the data logger can be grounded to a properly grounded metal pipe
or fixture.
A38-P Features:
Pipe attachment
ground kit
Clamps to ¾ in - 1 ¼
inch pipe
8 ft of #8 copper wire
Installation:
Attach clamp to
grounded pipe
Attach bar copper
wire end to iSIC
ground lug.
Potted Nut