9.4 Operating Menu Setup
9.4.1 Flow System Initial Setup
In order to conduct measurements, the ‘Basic Config’ setup must be
completed. If power goes off, this setup will remain in memory and
does not have to be re-entered.
Press any key except ‘EXIT’ on the keypad to enter the ‘Menu
Mode’.
When ‘Basic Configmenu is displayed on Line 1 of the display
press the ‘ENTER’ key to enter the options available under that
menu.
Sensor Serial #’ will appear on Line 2 of the display. Press the
ENTER’ key and the current sensor serial number entered in the
transmitter will be displayed on Line 4 of the display. The user can
then use the arrow keys ( and to scroll through the values and
and to change position) to enter the sensor band serial number
found on the sticker attached to the sensor band. Once all digits
are entered, press the ENTER’ key to save to memory. ‘→Sensor
Serial #’ will again be displayed on Line 2.
Next, press the key to scroll to the ‘Pipe Size’ menu on Line 2.
Press the ENTER’ key to access the options available under that
menu. Note: it is necessary to select only one of the following
options.
The first choice on the Pipe Size’ menu is ‘ID / Wall’ shown on
Line 3 of the display. If this value is known it can be entered here.
Press the ENTER’ key and the current saved inner diameter and
wall thickness will be shown on Line 4 of the display. Use the
arrow keys ( and to scroll through the values and and to
change character) to enter the pipe inner diameter. Units can be
inches or millimeters. Once all digits are entered, press the
ENTER’ key to save the value to memory.
The second choice on the Pipe Size’ menu is ‘Size/Sched’ (pipe
Size / Schedule). If this value is known, press the ENTER’ key.
Use the arrow keys to enter the values and then press the ENTER
key to save to memory.
The third choice on the Pipe Size’ menu is ‘OD / Wall’ displayed
on Line 3. Use the arrow keys ( and to scroll through the values
and and to change character) to enter the pipe outer
diameter. Units can be inches or millimeters. Once all digits are
entered, press the ENTER’ key to save the value to memory.
Note: Once a pipe size has been entered, re-entering the ‘Pipe
Size’ menu and selecting Size/Sched’ may cause errors in scaled
output values.