Instruction Manual

Copyright © 2006 CiDRA Corporation Page 9-19
20675-01 Rev 02
Level 1 Level 2 Level 3 Range Description
State Enable, Disable Enable Damping
Flow & GVF
Damping
Time
Constant (s)
0-600 seconds; default
is 3
Smoothes output due to
rapid changes in flow
State Enable, Disable Enable noise filtering
Flow & GVF
Noise Filter
Magnitude Low, High Amount of damping
State Enable, Disable Enable spike filtering
Length 2 - 60 readings
# of measurements to
validate
Output Config (cont)
Flow & GVF
Spike Filter
Percent or
Delta
0 – 100 %
% of full measurement
range that will define
the spike size to be
rejected