Table 5 Sensor to Transmitter Cable Terminal Connections
8.4.2.3.2 Armored cable installation
For those installations that will use armored sensor to transmitter
cable, install the P/N 20448-01 transmitter housing stiffener plate
(shipped with the armored cable) in the transmitter housing. Ensure
the plate is installed with the bent edge up and located in the front
portion of the transmitter housing. The stiffener plate is held in place
by the cable gland fittings.
Figure 37 Armored Cable Stiffener Plate Installation
Wire Pair # Transmitter
Terminal #
Function SENSOR CONNECTOR
PIN NUMBER
1 Wht/Blk 1 Sensor #1 Input 1 - Wht 13 - Blk
2 Wht/Blk 2 Sensor #2 Input 2 - Wht 14 - Blk
3 Wht/Blk 3 Sensor #3 Input 3 - Wht 15 - Blk
4 Wht/Blk 4 Sensor #4 Input 4 - Wht 16 - Blk
5 Wht/Blk 5 Sensor #5 Input 5 - Wht 17 - Blk
6 Wht/Blk 6 Sensor #6 Input 6 - Wht 18 - Blk
7 Wht/Blk 7 Sensor #7 Input 7 - Wht 19 - Blk
8 Wht/Blk 8 Sensor #8 Input 8 - Wht 20 - Blk
9 Wht/Blk 9 Spare – unused
--- ---
10 Wht/Blk 10 Wht – RS 485 Hi / Blk – RS485 Low 12 - Wht 24 - Blk
11 Wht/Blk 11 Wht – ‘-12 V’ / Blk – Gnd 9 - Wht 21 - Blk
12 Wht/Blk 12 Wht – ‘+12 V’ / Blk – Gnd 10 - Wht 22 - Blk