Section 2: Installation and Setup
2.2.4 Advanced Optical Alignment
Boresight Alignment Procedure
1. Display the boresight test pattern by pressing the T
EST key on the remote keypad or use the built-in keypad
and press the soft key that displays “Test” on the LCD display. Press the U
P ARROW KEY to cycle to
Boresight, then Enter.
2. Focus the image on cross-hair image I. Evaluate the focus on cross-hair image II and III. If all 3 images are
in focus, no further action is required. If boresight is required, continue to step 3.
3. See Figure 2-1 Cross-Hair Pattern to understand how the adjustment screws on the lens mount affect the
corresponding cross-hairs on the test pattern.
NOTE: The cap screw, (Figure 2-2 Screw Locations) may be under a plastic cap. Remove and retain
before adjusting.
4. Use a 5 mm hex driver to loosen 3 setscrews on the lens mount as shown in Figure 2-2 Screw Locations.
NOTE: The setscrews must be backed out several turns to avoid contact with the inner lens mount plate.
5. Fine tune the focus of cross-hair pattern I by adjusting the appropriate cap screw. Adjust until the cross-
hair image is in focus with minimal flare.
FIGURE 2-1 CROSS-HAIR PATTERN FIGURE 2-2 SCREW LOCATIONS