Datasheet
Datasheet CLM1000
Power measurement instrument
Alpenstraße 34
87700 Memmingen
E462753 Telefon (0 83 31) 83 71-0 Amtsgericht
Rev03 Telefax (0 83 31) 83 71-99 Memmingen
eMail: info@christ-elektronik.de HRB 9102
Internet:
http://www.christ-elektronik.de
Seite 2 / 2
Display und Handling
Display 128*64 display with different backlight colours
Operating elements 3 Buttons
Method of Measurement and Accuracy
Method of Measurement
Voltage directly measured at the consumer, current measured
by precision shunt
Measuring rate Approx. 1 second
Sampling rate Approx. 2000 Hz
Zero load detection If I < 0,002 Ampere current and power values are set to zero.
It P < 0,2 Watt power values and current are set to zero.
In this case the power factor is set to 1,00.
The type of resistance is set to ohm-resistance.
EEPROM-Memory All settings and consumption values are stored at line interrupt
/power off
Accuracy ±0,3%±3 digit of Display reading (PF > 0,3)
Voltage supply
Connector Interconnect socket, permanent current load 16A
Supply voltage 100 - 264 V
AC
, 47 - 63 Hz
Power consumption < 4VA
Dimensions and Environment
Size ABS plastic approx. 95*35*200 [mm] (W*H*L)
Weight approx. 490 g
Cable approx. 1,2m
Operating temperatur
0°C - 50°C, no dew allowed
Protection IP 50 (DIN EN 60529) (USB- Interface IP40)
Protective class II (prot. Isolation) DIN EN 61140
Measurement category CAT II
Changes may occur without notice!