3800 N. Mill Road Vineland, NJ 08360 Phone: 800.843.1794 Fax: 800.922.4361 ASSEMBLY INSTRUCTIONS FOR UNJACKETED FILTER REACTOR SYSTEMS Support Frame with Oversized Rugged S.S. Uprights and Heavy Duty PTFE Coated Aluminum Fittings Open Frame Design for Excellent Accessibility Mechanical Seal Stir Bearing PTFE Filter Base 20L Vessel Body Detachable Drain Valve Heavy Duty Lockable Castors www.cglifesciences.com technical-service@cglifesciences.
Read Entire Assembly Instructions Before You Begin. Familiarize Yourself with All of the Parts, and Pay Close Attention to All Notes and Highlights. Support frames are shipped via common carrier and require loading dock access with a fork lift or jack. If you do not have a loading dock, then a lift gate-equipped truck must be requested at the time of order.
Overhead Stirrer Motor Air Motor 1/4 HP Horizontal Motor 1/4 HP Vertical Motor Explosion Proof (XP) Motor The motor is installed on the support frame when shipped. Use the following instructions should you need to remove or adjust these components.
Stirrer Shaft and Agitator Assembly CG-2097 Stirrer Shaft CG-2095 Upper Agitator Hole through the Lower Portion of the Stirrer Shaft CG-2096 Lower Agitator Glass Filled PTFE Pin (MUST BE CUT TO LENGTH BEFORE INSTALLING) The stirrer shaft and agitators are assembled and installed when shipped. Use the following instructions should you need to remove or adjust these components.
Filter Base System Centering Ring Frit (Filter) *Note: The Bottom Impeller Should NOT Rub on the Filter Perforated Plate Filter Base 1. 2. 3. 4. 5. Place the o-ring in the beveled edge of the filter base. Place the perforated plate in the filter base, beveled edge facing up. Place the frit (filter) on top of the perforated plate. Fix the centering ring on top of the frit and make sure it is pushed down on all edges evenly. Wrap the lower outlet adapters with NPT end with PTFE tape.
BEFORE YOU START: The reactor is now ready for use. Before filling the reactor, make sure the 0-14mm drain valve is in the closed position. Once setup is complete, the reactor can be moved into position and connected to the circulator. Keep the wheels locked for added safety. Chemglass presumes some knowledge of this type of equipment on the part of the end user.
www.cglifesciences.com technical-service@cglifesciences.
Optional Components: Safety Shield Side Support Shelf Powder Addition Funnel www.cglifesciences.com Distillation Side Support Shelf Data Logger technical-service@cglifesciences.
CG-1978-P and CG-3498 Probes for Reactor Systems Installed in Hazardous/XP or ATEX environments. Chemglass PT100 and thermocouple probes that will be used in a hazardous/explosion proof area must be used with a Zener barrier. Zener barriers are used in control and instrumentation systems for the process of standardized signals, such as 20 mA or 10 V. Zener barriers contain intrinsically safe circuits that are to be used to drive intrinsically safe field devices with hazardous area.