Magnetic Hotplate-Stirrer with Timer User Manual OptiMag-ST Magnetic Hotplate-Stirrer with Timer Please read the User Manual carefully before use, and follow all operating and safety instructions! Technical specifications and outline are subject to change without prior notice.
Contents Service ............................................................................ 1 Warranty ......................................................................... 1 1 Safety Instructions .................................................. 2 2 Proper Use .............................................................. 3 3 Inspection................................................................ 4 3.1 Receiving Inspection ............................................ 4 3.2 Listing of Items ......
• Certification • Description of problem (i.e., hardware or software) • Methods and procedures adopted to resolve the problems • Your contact information Preface Welcome to the “(Hotplate) Timer Magnetic Stirrer User Manual”. Users should read this Manual carefully, follow the instructions and procedures, and be aware of all the cautions when using this instrument.
1 Safety Instructions substances or around water. • Gradually increase the speed, reduce the speed if: - Stirring bar breaks away due to high speed - The instrument is not running smoothly, or the container is not centered on the plate • Temperature must always be set to at least 50C lower than the flash point of the media used.
Do not use damaged components. Safe operation is only guaranteed with the accessories listed in the “Accessories” section. Accessories must be securely attached to the device and must be removed when not in use. Always disconnect the power before fitting accessories. • When using the external temperature sensor, the tip of the measuring sensor must be at least 5-10mm from vessel’s bottom and wall.
3 Inspection 4. Control 3.1 Receiving Inspection 4.1 Control elements Unpack the equipment carefully and check for any damage which may have arisen during transport. Please contact manufacturer/supplier for technical support. Note: If there is any apparent damage to the unit, do not attempt to operate. 3.
Figure 3 5
Items Function switch knob Turning the knob, switches the parameters setting interface, followed by Temperature setting, Speed setting, Timer setting, switching cycle. Pressing the knob turns the three setting options on/off. Parameter setting knob At the parameter setting interface, set the parameter by turning the knob. Pressing the knob turns on the initializer. LCD display LCD displays the real-time working state and all settings.
4.2 Display Probe Display when using external probe. Remote Display when using remote control. Rated value/Real value Display values in when heating or stirring or timer function is switched ON. Table 3 Note: If both heating and stirring functions have been started at the same time, heat display always has higher priority. If in this case speed is changed via the stirring knob, it displays stirring speed and reverts to temperature after 5 seconds.
• • • • • • • Set the target stirring speed and begin. Observe the stirring bar and LCD display. Set the target temperature and start heating. Observe the real temperature on LCD display. Set the timer. Timer ends, the heating function stops. Stops stirrer function. If these operations above are normal, the device is ready to operate. If these operations are not normal, the device may be damaged during transportation, please contact manufacturer/supplier for technical support.
6.1 Working with the PT1000 external temperature sensor following at this time: • • Turn off the power Ensure the external temperature sensor is positioned sample container correctly • Turn on the power, set the target temperature and switch on the heating function • If the unit is still not functioning correctly, please contact manufacturer/supplier 6.2 Residual heat warning(HOT) Figure 6 The PT1000 external temperature sensor is the manufacturer’s optional accessory.
7 Stirring 9 Remote Control The “stirring” function is switched on or off via rotating stirring knob. The speed is set on the knob (100 to 1500 rpm in steps of 10 rpm). When both the heating function and stirring are switched on and the above operations are completed, the LCD will shift to the speed value and revert back to the temperature value in about 5 seconds. The last speed value set is displayed.
• Mode of transmission: Fully Duplex. 1 start bit; 7 character bits; 1 parity bit [straight (even)); 1 stop bit • Transmission speed: 9600 bit/s Note: Forbid to insert or remove the communication line when switch on! RS232C 10 Faults • Instruments can’t be powered ON - Check whether the power line is unplugged - Check whether the fuse is broken or loose • Fault in power during self-test - Switch OFF the unit, then switch ON and reset the instruments to factory default setting.
• Before using other method for cleaning or decontamination, the user must ascertain with the manufacturer that this method will not harm the instrument. Wear the proper protective gloves during cleaning of the instrument. 12 Associated Regulations Standards Note: • Electronic equipment cannot cleaned with a detergent cleaner. • Repairing equipment must be cleaned and avoid the contamination of dangerous substances, put the equipment into original package box and returned.
13 Specifications Temperature display accuracy [°C] ±0.1 The safety temperature of the hotplate [℃] 360 Temperature sensor in medium PT1000 ±0.2 Items Digital model Voltage [VAC] 200-240 / 100-120 Frequency [Hz] 50 / 60 Power [W] 550 Stirring point position quantity 1 Max. stirring quantity (H2O) [L] 20 Control accuracy of heating temperature with temperature sensor [℃] Max.