DESCRIPTION OF FUNCTIONS
The following is a description of the standard features of the AWS product line.
OPERATING MODE
Peak
Displays and retains the maximum torque exerted by the wrench, as occurs
when operating the wrench in the tightening direction. The Peak Mode is used
for all power tools and some dial wrenches.
1st Peak
Detects the “first peak” of torque experienced by the wrench, capturing the
initial torque as occurs when the torque wrench cams over. First Peak is used
primarily for Click torque wrenches and cam over screwdrivers.
Track
Displays torque as it is being applied to the transducer. Track mode is used
primarily for verifying calibration of the unit.
ENGINEERING UNITS
Shows the current engineering units. Press the key to cycle through the eight
possible choices: Kgf m, KgfCm, gfCm, cNm, Nm, FT LB, IN LB, IN OZ.
FULL SCALE
This screen shows the Full-Scale value of the Torque Shaft. This is not a field
adjustable value.
LOW LIMIT
Use the low limit setting as a means of visually flagging the operator when a
reading fails to reach a desired minimum value. A small down arrow will
appear on the screen if a peak is captured below the limit setting.
HIGH LIMIT
Use the high limit setting as a means of visually flagging the operator when a
reading falls over a desired maximum value. High limits are set in the identical
way as low limits. Please refer to the Low Limit section for details.
NOTE ON LIMITS: The green LED on the front of the display will flash when a
peak is captured that falls within the limit setting.