3.0 OPERATING PROCEDURES
3.1 Mounting Arrangements
TS2 Series sensors can be mounted in different ways:
 
 
M3 Screws
M4 Screws
3.2 Connector Assignment
NOTE: To assure immunity to random noise, the measuring head must be installed

be connected to the metal housing of the connecting plug.
Pin assignment viewed from outside:
1 - Vcc + 15 to 24 V DC regulated
2 - Ground
3 - Ground
4 - Signal output -
5 - Signal output +
1
3
Shield
2
5
4
Female socket M9
– 7 –
3.3 Options
Code DC Analog output 0 - 10 V DC
Code MA Analog output 4 - 20 mA
3.4 Initial Setup
 
 
of the 5-pin connector located on the rear side of the sensor is described in
section 3.2.
 
signicantly from the factory calibration material, you need to carry out
ZERO Adjustment and GAIN Adjustment as described in sections 4.1 and
4.2 before starting measurement.
 
 
the material path symbol on the front of the sensor.
1
2
3
4
5
Strand assignment of cable EK0620:
1 (white) Vcc + 15 to 24 V DC regulated
2 (black) Ground
3 (brown) Ground
4 (blue) Signal output –
5 (grey) Signal output +