The TS2 Series of universal tension sensor accurately measure the running tension
of wires, cables, yarns, bers and similar process material.
2.1 Delivery
2.2 Unpacking
Unpack the instrument and inspect it for any shipping damage. Notices of defect
must be led immediately, in writing, at the latest within 10 days on receipt of the
goods.
2.3 Model and Serial Numbers
clylindrical portion of the sensor.
TS2-1K + ASY = TS2-1K-ASY
Model with
tension range
Code for guide rollers
(if not standard)
Complete Model No.
– 11 –
5.0 SPECIFICATIONS
5.1 General Specifications
Calibration According to factory procedure
Accuracy For 10% to 100% of range: ± 1% Full Scale
Other calibration material: ± 3% Full Scale or
better
Overload Protection 100% of range
Measuring Principle Strain gauge bridge
Meas. Roller Deflection 0.5 mm, max.
Natural Frequency of
Measuring Spring Approx. 500 Hz, depending on tension range
Signal Processing Analog
Output Signal 0 - 1 V DC, impedance: > 5 kOhm (standard)
Option Code DC 0 - 10 V DC
Damping (fg) Standard: approx. 30 Hz (other values on
request)
Temperature Coefficient Zero point: less than ± 0.05% FS* / °C
Temperature Range 10 – 45 °C
Air Humidity 85% RH, max.
Power Supply 15 to 24 V DC (21mA) (regulated)
Option code A3: max. 50 mA
Housing Aluminium